Special Survey Instructions - RSS Experiment

January 23, 2002

General Guidance for Surveys in Hall C During RSS Experiment

The purpose of this document is to clarify the requirements for radiation surveys in Hall C when the polarizing target magnet is energized. The attached photos should help reinforce the description that follows.

Specific Requirements:

  1. Only the Teletector 6112B "yellow jacket" instrument (or alternative approved by RCG) may be used for this purpose.
  2. Do not attempt to obtain a contact dose rate on the target chamber or on beamline within 1 foot of the chamber.
  3. This survey is acceptable for a "full survey" of the hall. However, if the Whole Body dose rate at the target chamber is greater than 2 mR/hr, the RCG shall be contacted prior to allowing direct, hands-on access to the target chamber (i.e. for Restricted Access).

General Practice:

*Remember to notify the Hall C shift leader of the entry when you get to the Counting House.


General area survey under pivot area
Typical dose rate ~ 0.2 mR/hr


Survey around HMS deck and pivot from below
Typical dose rate ~0.4 mR/hr


From the target/cryo control platform, extend probe through fence
for Whole Body survey of upstream beamline and target chamber
Typical dose rate ~ 0.4 mR/hr


From the upper target platform, survey entire upper platform area. This can be done without passing under the cables from the target to the rack.
Typical dose rate ~0.2 mR/hr


From the floor downstream of the target, do a Whole Body survey
of the exit beamline and the SOS side of the chamber.
Typical dose rate here is normally < 1 mR/hr