Privacy and Security Notice
Main INDEX
Monthly INDEX
PREV
NEXT
User name R. Michaels
Log entry time 11:17:41 on August 8,2007
Entry number 208775
This entry is a followup to: 208731
keyword=NaI gate and 5 new 1881M
Update on the NaI gate: The delay needs to be 250 nsec.
I've made this gate so I have better measurements, also this
delay takes into account the need to put the analog signals about
25 nsec after the start of the gate.
Five units of 1881M ADC from BNL were added to ROC4 slots 16-20.
Fastbus test code showed they were ok. The pedestals look fairly
normal except that some channels are rather noisy (I doubt it can
be blamed on the ADC, though).
Another problem at present is that the gate width is a maximum of
50 nsec. I don't have any other "adjuster knob" that can make a
wider gate, at the moment.