NEXT
User name a-onl
Log entry time 04:47:22 on May 16, 2008
Entry number 238196
keyword=Start_Run_1154,Run_type=Calibration,target_type=Loop 3 LH2 6 cm,beam_e0=1.192,beam_raster=on,leftarm_p=0.6286,leftarm_theta=57.5,leftarm_colli=out,bbarm_i=700,bbarm_theta=30.25,bbarm_zoff=,comment_text=checking FPP HV=1875 V
Run Timestamp: Fri May 16 04:46:38 EDT 2008
Beginning-of-run EPICS data for Run Number 1154 at Fri May 16 04:46:54 EDT 2008
Run_type=Calibration
target_type=Loop 3 LH2 6 cm
beam_e0=1.192
beam_raster=on
leftarm_p=0.6286
leftarm_theta=57.5
leftarm_colli=out
bbarm_i=700
bbarm_theta=30.25
bbarm_zoff=
comment_text=checking FPP HV=1875 V
; BBite+HRS DAQ on adaql2/a-onl account
; Prescale factors (integer), downloaded each run at prestart.
; Do NOT put spaces in string, NOR add 2nd string. (REALLY!!!)
; Change this file only in-between runs.
; Maxima: (2^24)-1 = 16777215 for T1-T4 and 65535 for T5-T8
; The default factor is essentially infinite.
; T1 = BigBite Calorimeter
; T2 = undefined
; T3 = L-HRS Singles (S1.and.S2)
; T4 = L-HRS Efficiency (S1.or.S2)
; T5 = T1.AND.T3 (BigBite LHRS Coincidence)
; T6 = undefined
; T7 = BigBite Cosmics
; T8 = 1024 Hz pulser
; String should be one line, no spaces, ending in carriage return
ps1=60000,ps2=16777215,ps3=1,ps4=1000,ps5=1,ps6=65535,ps7=1,ps8=100
Left Arm Q1 momentum : 0.628529
Left Arm Q1 power supply : 472.915
Left Arm Q1 gauss meter : 0.04056
Left Arm Q2 power supply current : 270.7
Left Arm Q2 momentum : 0.628657
Left Arm Q2 gauss meter : -0.10722
Left Arm Q2 power supply polarity : enabled
Left Arm D1 power supply current : 208.642
Left Arm D1 momentum : 0.628597
Left Arm D1 NMR : 0.264705
Left Arm Q3 power supply current : 250.259
Left Arm Q3 momentum : 0.628606
Left Arm Q3 gauss meter : -0.12689
Left Arm Q3 power supply polarity : enabled
Left Arm B Field Hall Probe (D-G) : 0.2251
-----
Right Arm Q1 momentum : 0.00339873
Right Arm Q1 power supply : 0
Right Arm Q1 gauss meter : -7.28e-05
Right Arm Q2 power supply current : -0.000118317
Right Arm Q2 momentum : -2.7492e-07
Right Arm Q2 gauss meter : -0.0002824
Right Arm Q2 power supply polarity : enabled
Right Arm D1 power supply current : 0.0269629
Right Arm D1 momentum : 0.00442046
Right Arm D1 NMR : 0.30125
Right Arm Q3 power supply current : 6.35172e-05
Right Arm Q3 momentum : 0.000257472
Right Arm Q3 gauss meter : -0.0002532
Right Arm Q3 power supply polarity : enabled
Right Arm B Field Hall Probe (D-G) : -0.001626
-----
BigBite magnet current setpoint : 700
BigBite magnet current readback : 696.619
BigBite Gaussmeter readback : 0.5472
-----
Left Arm Top VDC Current : -2.14844
Left Arm Bottom VDC Current : 4.10156
Left Arm Top VDC voltage : 3.98535
Left Arm Bottom VDC voltage : 3.99121
-----
Right Arm Top VDC Current : -1.12305
Right Arm Bottom VDC Current : -1.07422
Right Arm Top VDC voltage : -0.0112305
Right Arm Bottom VDC voltage : 0.00146484
-----
Average current of the two cavities : 3.68254
Current on Unser monitor : 17.103
Current on Upstream bcm : 3.61812
Current on Downstream bcm : 3.74696
Upstream Cavity Coefficient : 42.6
Downstream Cavity Coefficient : 43.8
Current on IBC0L02 (source) : 98.9457
Hall C Current : ibcm1 --- Invalid channel name
Hall A momentum (Tiefenbach) (HALLA:p) : 1192.27
Hall A momentum orbit correction (dpp) : -0.000244091
Hall A setup momentum (MMSHLAE) : 1193.56
Hall A energy (halla_MeV) : 1192.27
Hall C energy (hallc_MeV) : 5710.51
Delta(E)/E width (MV_AI_DXWID) : 1.49557
MBSY1c magnet current (amp) : 49.4616
MBSY1c Bdl : 297839
Arc1 dp/p : 0
Arc2 dp/p : 0
Raster X width : 1.5
Raster Y width : 2
-----
Left Arm Collimator : 0.871436
HacL_HPE1313A:0_Ch58 : -0.303223
HacL_HPE1313A:0_Ch55 : 23.2422
HacL_HPE1313A:0_Ch54 : -0.354004
Left Arm Angle Floor Angle : 57.5
Left Arm Angle Vernier : 31.1
Left Arm Scattering Angle : 57.5003
-----
Right Arm Arm Collimator : -0.00864258
HacR_HPE1313A:0_Ch58 : -0.574707
HacR_HPE1313A:0_Ch55 : 21.9238
HacR_HPE1313A:0_Ch54 : -0.842285
Right Arm Angle Floor Angle : 105
Right Arm Angle Vernier : -34.4
Right Arm Scattering Angle : 104.999
-----
IPM1C12.XPOS : 0.915224
IPM1C12.YPOS : 0.49034
IPM1H04A.XPOS : 1.13952
IPM1H04A.YPOS : 0.863843
IPM1H04B.XPOS : 0.898417
IPM1H04B.YPOS : 1.30146
Survey X offset bpmA : -0.63
Survey Y offset bpmA : -0.9
Survey X offset bpmB : 0.77
Survey Y offset bpmB : -1
-----
Wien angle : -33.7301
Helicity Mode ON/OFF Random/Toggle : OFF
Wien field : 22.411
Laser attenuation Hall A : 87
Laser power Hall A : 100
Half-wave plate (off=out) : IN
rotating waveplate : 0
Beam Half-wave plate (off=out) : IN
-----
G0 Helicity Settle Set : 500 usec
G0 Helicity Delay Set : No Delay
G0 Helicity Pattern Set : Quartet
G0 Helicity Cycle Rate Set : 30 Hz
G0 Helicity Settle Read : 500 usec
G0 Helicity Delay Read : No Delay
G0 Helicity Pattern : Quartet
G0 Helicity Cycle Rate Read : 30 Hz
-----
Target : L3 6cm
fan frequency loop 1 : 0
fan frequency loop 2 : 0
fan frequency loop 3 : 36
Encoder position : 2.06493e+07
Target type : 2.06493e+07
temperature loop 1(1) : 109.55
temperature loop 1(2) : 109.05
temperature loop 1(3) : 115.15
Heat exchange loop 1 H2 : 101.9
temperature loop 3(1) : 20
temperature loop 3(2) : 19.99
temperature loop 3(3) : 29.99
Heat exchange loop 3 H2 : 20
pressure loop He : 40.8157
pressure loop H2 before : 29.9687
pressure loop H2 return : 30.2449
HPH He loop 1 : 0.114237
HPH H2 loop 3 : 83.1576
----
SLI data valid? : No
SLI horiz size mm : 0
SLI Energy spread *10^5 : 0
SLI x position : 2.32547
SLI vert. size mm : 0
SLI y position : 2.35117
-----
FPP High Voltages and Currents at Fri May 16 04:47:06 EDT 2008
MF7b1469_7gn0gen.F 1875
MF7b1469_7gn0gen.E 1.6
MF7b1469_7gn0ch0property.F 1875
MF7b1469_7gn0ch0property.E 0.04
MF7b1469_7gn0ch1property.F 1875
MF7b1469_7gn0ch1property.E 0.12
MF7b1469_7gn0ch2property.F 1875
MF7b1469_7gn0ch2property.E 0.04
MF7b1469_7gn0ch3property.F 1875
MF7b1469_7gn0ch3property.E 0.06
MF7b1469_7gn0ch4property.F 1875
MF7b1469_7gn0ch4property.E 0.09
MF7b1469_7gn1gen.F 1875.2
MF7b1469_7gn1gen.E 1.1
MF7b1469_7gn1ch0property.F 1875.2
MF7b1469_7gn1ch0property.E 0.14
MF7b1469_7gn1ch1property.F 1875.2
MF7b1469_7gn1ch1property.E 0.1
MF7b1469_7gn1ch2property.F 1875.2
MF7b1469_7gn1ch2property.E 0.03
MF7b1469_7gn1ch3property.F 1875.2
MF7b1469_7gn1ch3property.E 0.06
MF7b1469_7gn1ch4property.F 1875.2
MF7b1469_7gn1ch4property.E 0.24
MF7b1469_8gn0gen.F 1874.6
MF7b1469_8gn0gen.E 1.2
MF7b1469_8gn0ch0property.F 1874.6
MF7b1469_8gn0ch0property.E 0.05
MF7b1469_8gn0ch1property.F 1874.6
MF7b1469_8gn0ch1property.E 0.05
MF7b1469_8gn0ch2property.F 1874.6
MF7b1469_8gn0ch2property.E 0.05
MF7b1469_8gn0ch3property.F 1874.6
MF7b1469_8gn0ch3property.E 0.04
MF7b1469_8gn0ch4property.F 1874.6
MF7b1469_8gn0ch4property.E 0.03
MF7b1469_8gn1gen.F 1874.2
MF7b1469_8gn1gen.E 1.1
MF7b1469_8gn1ch0property.F 1874.2
MF7b1469_8gn1ch0property.E 0.01
MF7b1469_8gn1ch1property.F 1874.2
MF7b1469_8gn1ch1property.E 0.14
MF7b1469_8gn1ch2property.F 1874.2
MF7b1469_8gn1ch2property.E 0.05
MF7b1469_8gn1ch3property.F 1874.2
MF7b1469_8gn1ch3property.E 0.05
MF7b1469_8gn1ch4property.F 1874.2
MF7b1469_8gn1ch4property.E 0.19
MF7b1469_8gn2gen.F 23.4
MF7b1469_8gn2gen.E 0
MF7b1469_8gn2ch0property.F 0
MF7b1469_8gn2ch0property.E 0
FPP Chamber Threshold voltages at Fri May 16 04:47:07 EDT 2008
HrsH_FPP_TH1 3.91016
HrsH_FPP_TH2 4.02637
HrsH_FPP_TH3 6.96094
HrsH_FPP_TH4 7.07178
FPP Wire Chambers Low Voltage at Fri May 16 04:47:08 EDT 2008
HrsH_FPP_LV11P 4.91992
HrsH_FPP_LV11M -5.34277
HrsH_FPP_LV12MA -5.25439
HrsH_FPP_LV12MB -5.58691
HrsH_FPP_LV21P 5.08105
HrsH_FPP_LV21M -4.54688
HrsH_FPP_LV22MA -4.7124
HrsH_FPP_LV22MB -4.73193
HrsH_FPP_LV31P 4.77686
HrsH_FPP_LV31M -5.10498
HrsH_FPP_LV32MA -2.80078
HrsH_FPP_LV32MB -1.51514
HrsH_FPP_LV42P 4.76562
HrsH_FPP_LV42MA -4.7168
HrsH_FPP_LV42MB -4.5752
HrsH_FPP_LV41P 4.73193
HrsH_FPP_LV41MA -4.77002
HrsH_FPP_LV41MB -4.60498
FPP Level Shifters at Fri May 16 04:47:08 EDT 2008
HacH_DTC_HPE1313A:0_Ch48 5.07129
HacH_DTC_HPE1313A:0_Ch50 -5.11377
FPP Gas Flow at Fri May 16 04:47:08 EDT 2008
HrsH_GAS_rFPPc1 4.36913
HrsH_GAS_rFPPc2 2.25102
HrsH_GAS_rFPPc3 3.87212
HrsH_GAS_rFPPc4 5.58964
FPP Carbon Door Status at Fri May 16 04:47:09 EDT 2008
HacH_DTC_VMIC1182:0_Ch13 enabled
HacH_DTC_VMIC1182:0_Ch17 enabled
HacH_DTC_VMIC1182:0_Ch9 enabled
HacH_DTC_VMIC1182:0_Ch5 enabled
Left S2 HV
S2 Slot0 ch0 : -2311
S2 Slot0 ch1 : -2212.3
S2 Slot0 ch2 : -2165.7
S2 Slot0 ch3 : -2231.2
S2 Slot0 ch4 : -1793.1
S2 Slot0 ch5 : -2166.5
S2 Slot0 ch6 : -1940.7
S2 Slot0 ch7 : -1876.9
S2 Slot0 ch8 : -2030.4
S2 Slot0 ch9 : -1889.6
S2 Slot0 ch10 : -2094.7
S2 Slot0 ch11 : -1867.9
-----
Left S1 HV
S1 Slot1 ch0 : -1785.2
S1 Slot1 ch1 : -1809.8
S1 Slot1 ch2 : -1790.2
S1 Slot1 ch3 : -1817.7
S1 Slot1 ch4 : -1699.9
S1 Slot1 ch5 : -1855.4
S1 Slot1 ch6 : -1915.4
S1 Slot1 ch7 : -1700.9
S1 Slot1 ch8 : -1961.1
S1 Slot1 ch9 : -1746.3
S1 Slot1 ch10 : -1798.4
S1 Slot1 ch11 : -1707.6
-----
Left Cerenkov HV
GCER Slot2 ch0 : -1543.5
GCER Slot2 ch1 : -1462.1
GCER Slot2 ch2 : -1528
GCER Slot2 ch3 : -1453
GCER Slot2 ch4 : -1798.1
GCER Slot2 ch5 : -1983.6
GCER Slot2 ch6 : -1750.5
GCER Slot2 ch7 : -1751.9
GCER Slot2 ch8 : -2009.9
GCER Slot2 ch9 : -1588.6
GCER Slot2 ch10 : -15
GCER Slot2 ch11 : -23.6
-----
Left Pion Rejector Layer 1 HV
PRL1 Slot3 ch0 : -860.4
PRL1 Slot3 ch1 : -784.5
PRL1 Slot3 ch2 : -905.8
PRL1 Slot3 ch3 : -980.5
PRL1 Slot3 ch4 : -731.1
PRL1 Slot3 ch5 : -811
PRL1 Slot3 ch6 : -1081.3
PRL1 Slot3 ch7 : -990.8
PRL1 Slot3 ch8 : -810.5
PRL1 Slot3 ch9 : -790.9
PRL1 Slot3 ch10 : -829.5
PRL1 Slot3 ch11 : -785.3
PRL1 Slot4 ch0 : -870.1
PRL1 Slot4 ch1 : -900.5
PRL1 Slot4 ch2 : -750.8
PRL1 Slot4 ch3 : -895.1
PRL1 Slot4 ch4 : -921.7
PRL1 Slot4 ch5 : -821.2
PRL1 Slot4 ch6 : -836
PRL1 Slot4 ch7 : -826.4
PRL1 Slot4 ch8 : -1190.3
PRL1 Slot4 ch9 : -991.4
PRL1 Slot4 ch10 : -1030.9
PRL1 Slot4 ch11 : -861.4
PRL1 Slot5 ch0 : -1041.1
PRL1 Slot5 ch1 : -871
PRL1 Slot5 ch2 : -916.2
PRL1 Slot5 ch3 : -735.8
PRL1 Slot5 ch4 : -890.7
PRL1 Slot5 ch5 : -810.3
PRL1 Slot5 ch6 : -821.1
PRL1 Slot5 ch7 : -895.6
PRL1 Slot5 ch8 : -741.3
PRL1 Slot5 ch9 : -841.4
PRL1 Slot5 ch10 : -16.5
PRL1 Slot5 ch11 : -25.6
-----
Left Pion Rejector Layer 2 HV
PRL2 Slot9 ch0 : -920.5
PRL2 Slot9 ch1 : -881.3
PRL2 Slot9 ch2 : -800.1
PRL2 Slot9 ch3 : -869.9
PRL2 Slot9 ch4 : -810.9
PRL2 Slot9 ch5 : -941.5
PRL2 Slot9 ch6 : -851.5
PRL2 Slot9 ch7 : -890.6
PRL2 Slot9 ch8 : -811.4
PRL2 Slot9 ch9 : -780.6
PRL2 Slot9 ch10 : -821
PRL2 Slot9 ch11 : -851.4
PRL2 Slot10 ch0 : -755.7
PRL2 Slot10 ch1 : -812.4
PRL2 Slot10 ch2 : -871
PRL2 Slot10 ch3 : -900.8
PRL2 Slot10 ch4 : -881.4
PRL2 Slot10 ch5 : -871
PRL2 Slot10 ch6 : -815.2
PRL2 Slot10 ch7 : -824.8
PRL2 Slot10 ch8 : -760.3
PRL2 Slot10 ch9 : -715.7
PRL2 Slot10 ch10 : -911
PRL2 Slot10 ch11 : -781.3
PRL2 Slot11 ch0 : -1000.7
PRL2 Slot11 ch1 : -900.4
PRL2 Slot11 ch2 : -850.6
PRL2 Slot11 ch3 : -1079.9
PRL2 Slot11 ch4 : -1050.2
PRL2 Slot11 ch5 : -800
PRL2 Slot11 ch6 : -860.4
PRL2 Slot11 ch7 : -1031.3
PRL2 Slot11 ch8 : -911.3
PRL2 Slot11 ch9 : -880.3
PRL2 Slot11 ch10 : -12.9
PRL2 Slot11 ch11 : -24.7
-----
# Map BB_DE Measured Voltage
1 : -1316 -1496
2 : -1420 -1333
3 : -1301 -1467
4 : -1632 -1618
5 : -1362 -1251
6 : -1444 -1444
7 : -1284 -1302
8 : -1345 -1601
9 : -1174 -1651
10 : -1567 -1301
11 : -1481 -1528
12 : -1281 -1564
13 : -1583 -1549
14 : -1167 -1550
15 : -1380 -1573
16 : -1403 -1301
17 : -1161 -1600
18 : -1431 -1400
19 : -1514 -1521
20 : -1369 -1546
21 : -1559 -1414
22 : -1494 -1619
23 : -1737 -1414
24 : -1558 -1284
# Map BB_DE Measured Current
1 : -135 -166
2 : -152 -147
3 : -138 -167
4 : -173 -182
5 : -143 -133
6 : -158 -162
7 : -141 -141
8 : -143 -172
9 : -127 -179
10 : -165 -143
11 : -156 -163
12 : -140 -176
13 : -171 -163
14 : -131 -170
15 : -147 -173
16 : -150 -137
17 : -123 -169
18 : -164 -153
19 : -168 -157
20 : -155 -170
21 : -167 -149
22 : -166 -157
23 : -196 -148
24 : -168 -138
# Map BB_E Measured Voltage
1 : -1456 -1479
2 : -1649 -1283
3 : -1298 -1334
4 : -1206 -1193
5 : -1330 -1555
6 : -1410 -1566
7 : -1550 -1284
8 : -1184 -1484
9 : -1287 -1514
10 : -1251 -1221
11 : -1572 -1472
12 : -1229 -1544
13 : -1632 -1361
14 : -1240 -1420
15 : -1565 -1254
16 : -1244 -1555
17 : -1488 -1389
18 : -1133 -1206
19 : -1345 -1540
20 : -1493 -1460
21 : -1440 -1454
22 : -1480 -1155
23 : -1391 -1227
24 : -1143 -3
# Map BB_E Measured Current
1 : -149 -159
2 : -180 -136
3 : -136 -146
4 : -134 -130
5 : -145 -169
6 : -150 -172
7 : -165 -141
8 : -131 -161
9 : -136 -157
10 : -133 -131
11 : -165 -153
12 : -136 -163
13 : -180 -136
14 : -127 -146
15 : -173 -131
16 : -134 -148
17 : -164 -152
18 : -127 -124
19 : -141 -163
20 : -159 -159
21 : -151 -150
22 : -156 -118
23 : -149 -129
24 : -119 -2
# Map BB_SC_R Measured Voltage
1 : -1176
2 : -1154
3 : -1174
4 : -1159
5 : -1190
6 : -1201
7 : -1267
8 : -1126
9 : -1196
10 : -1274
11 : -1306
12 : -1176
13 : -1293
# Map BB_SC_R Measured Current
1 : -442
2 : -437
3 : -444
4 : -439
5 : -448
6 : -456
7 : -481
8 : -427
9 : -453
10 : -483
11 : -495
12 : -446
13 : -489
# Map BB_DC Measured Voltage
1 : -0 -0 -0 -0
2 : 0 -0 0 -0
3 : -0 1 -0 0
# Map BB_DC Measured Current
1 : -0 -0 0 -0
2 : -0 -0 -0 -0
3 : -0 -0 -0 -0
# Map BB_PS_R Measured Voltage
1 : -883
2 : -842
3 : -845
4 : -853
5 : -858
6 : -848
7 : -829
8 : -844
9 : -895
10 : -897
11 : -842
12 : -856
13 : -860
14 : -849
15 : -817
16 : -917
17 : -787
18 : -807
19 : -824
20 : -810
21 : -815
22 : -796
23 : -815
24 : -800
25 : -765
26 : -848
27 : -850
# Map BB_PS_R Measured Current
1 : -226
2 : -215
3 : -212
4 : -219
5 : -219
6 : -217
7 : -213
8 : -216
9 : -229
10 : -232
11 : -210
12 : -219
13 : -221
14 : -213
15 : -212
16 : -234
17 : -202
18 : -201
19 : -213
20 : -207
21 : -211
22 : -204
23 : -206
24 : -205
25 : -196
26 : -213
27 : -221
# Map BB_SC_L Measured Voltage
1 : -1233
2 : -1271
3 : -1215
4 : -1140
5 : -1162
6 : -1268
7 : -1316
8 : -1122
9 : -1201
10 : -1243
11 : -1317
12 : -1230
13 : -1242
# Map BB_SC_L Measured Current
1 : -448
2 : -426
3 : -454
4 : -410
5 : -395
6 : -458
7 : -463
8 : -404
9 : -420
10 : -480
11 : -481
12 : -474
13 : -473
# Map BB_PS_L Measured Voltage
1 : -895
2 : -885
3 : -816
4 : -888
5 : -859
6 : -867
7 : -856
8 : -804
9 : -869
10 : -894
11 : -861
12 : -865
13 : -864
14 : -858
15 : -885
16 : -861
17 : -869
18 : -914
19 : -844
20 : -864
21 : -847
22 : -845
23 : -840
24 : -862
25 : -898
26 : -854
27 : -887
# Map BB_PS_L Measured Current
1 : -232
2 : -227
3 : -207
4 : -228
5 : -220
6 : -223
7 : -217
8 : -206
9 : -223
10 : -226
11 : -222
12 : -222
13 : -222
14 : -220
15 : -227
16 : -221
17 : -225
18 : -231
19 : -218
20 : -222
21 : -217
22 : -217
23 : -215
24 : -221
25 : -226
26 : -219
27 : -228
# Map BB_SH Measured Voltage
1 : -1316 -1420 -1301 -1632 -1585 -1721 -1322
2 : -1362 -1444 -1284 -1469 -1329 -1252 -1517
3 : -1345 -1174 -1567 -1481 -1456 -1552 -1496
4 : -1281 -1583 -1167 -1586 -1501 -1801 -1528
5 : -1380 -1403 -1161 -1431 -1401 -1376 -1484
6 : -1514 -1369 -1559 -1702 -1666 -1415 -1277
7 : -1494 -1737 -1558 -1496 -1617 -1417 -1577
8 : -1333 -1467 -1618 -1051 -1586 -1683 -1426
9 : -1251 -1444 -1302 -1601 -1459 -1301 -1651
10 : -1651 -1301 -1528 -1171 -1232 -1276 -1494
11 : -1564 -1549 -1550 -1573 -1351 -1400 -1300
12 : -1301 -1600 -1400 -1601 -1500 -1598 -1666
13 : -1521 -1546 -1414 -1619 -1216 -1526 -1557
14 : -1414 -1284 -1456 -1322 -1503 -1530 -1534
15 : -1649 -1298 -1206 -1330 -1519 -1412 -1461
16 : -1410 -1550 -1184 -1591 -1386 -1538 -1212
17 : -1287 -1251 -1572 -1229 -1418 -1481 -1523
18 : -1632 -1240 -1565 -1189 -1627 -1402 -1342
19 : -1244 -1488 -1133 -1345 -1343 -1603 -1316
20 : -1493 -1440 -1480 -1279 -1671 -1366 -1463
21 : -1391 -1143 -1479 -1283 -1571 -1402 -1234
22 : -1334 -1193 -1555 -1377 -1566 -1518 -1421
23 : -1566 -1284 -1484 -1514 -1601 -1180 -1322
24 : -1221 -1472 -1544 -1464 -1281 -1471 -1346
25 : -1361 -1420 -1254 -1555 -1327 -1304 -1510
26 : -1389 -1206 -1540 -1281 -1263 -1522 -1417
27 : -1460 -1454 -1155 -1227 -1261 -1420 -1425
# Map BB_SH Measured Current
1 : -135 -152 -138 -173 -170 -184 -140
2 : -143 -158 -141 -155 -141 -132 -166
3 : -143 -127 -165 -156 -152 -164 -158
4 : -140 -171 -131 -167 -160 -183 -163
5 : -147 -150 -123 -164 -134 -145 -153
6 : -168 -155 -167 -179 -175 -145 -138
7 : -166 -196 -168 -166 -173 -144 -168
8 : -147 -167 -182 -114 -166 -186 -152
9 : -133 -162 -141 -172 -159 -140 -180
10 : -179 -143 -163 -129 -134 -132 -161
11 : -176 -161 -174 -171 -147 -147 -142
12 : -138 -167 -147 -169 -162 -171 -178
13 : -163 -172 -149 -149 -128 -164 -163
14 : -151 -137 -155 -141 -160 -168 -162
15 : -179 -138 -131 -145 -158 -159 -154
16 : -151 -164 -123 -170 -146 -163 -128
17 : -135 -135 -166 -127 -147 -158 -162
18 : -174 -130 -170 -126 -179 -152 -141
19 : -136 -165 -124 -142 -143 -170 -134
20 : -163 -149 -158 -142 -175 -146 -151
21 : -148 -122 -157 -136 -168 -147 -132
22 : -146 -130 -166 -140 -169 -162 -157
23 : -173 -140 -161 -164 -165 -130 -139
24 : -129 -155 -164 -161 -135 -157 -146
25 : -136 -146 -131 -148 -140 -139 -161
26 : -152 -124 -163 -138 -138 -162 -150
27 : -159 -150 -118 -129 -131 -150 -156
# Map BB_TESTMAP Measured Voltage
1 : -1460 -1454 -1155 -1227 -1261 -1420 -1425
2 : -1389 -1206 -1540 -1281 -1263 -1522 -1417
3 : -1361 -1420 -1254 -1555 -1327 -1304 -1510
4 : -1221 -1472 -1544 -1464 -1281 -1471 -1346
5 : -1566 -1284 -1484 -1514 -1601 -1180 -1322
6 : -1334 -1193 -1555 -1377 -1566 -1518 -1421
7 : -1391 -1143 -1479 -1283 -1571 -1402 -1234
8 : -1493 -1440 -1480 -1279 -1671 -1366 -1463
9 : -1244 -1488 -1133 -1345 -1343 -1603 -1316
10 : -1632 -1240 -1565 -1189 -1627 -1402 -1342
11 : -1287 -1251 -1572 -1229 -1418 -1481 -1523
12 : -1410 -1550 -1184 -1591 -1386 -1538 -1212
13 : -1649 -1298 -1206 -1330 -1519 -1412 -1461
14 : -1414 -1284 -1456 -1322 -1503 -1530 -1534
15 : -1521 -1546 -1414 -1619 -1216 -1526 -1557
16 : -1301 -1600 -1400 -1601 -1500 -1598 -1666
17 : -1564 -1549 -1550 -1573 -1351 -1400 -1300
18 : -1651 -1301 -1528 -1171 -1232 -1276 -1494
19 : -1251 -1444 -1302 -1601 -1459 -1301 -1651
20 : -1333 -1467 -1618 -1051 -1586 -1683 -1426
21 : -1494 -1737 -1558 -1496 -1617 -1417 -1577
22 : -1514 -1369 -1559 -1702 -1666 -1415 -1277
23 : -1380 -1403 -1161 -1431 -1401 -1376 -1484
24 : -1281 -1583 -1167 -1586 -1501 -1801 -1528
25 : -1345 -1174 -1567 -1481 -1456 -1552 -1496
26 : -1362 -1444 -1284 -1469 -1329 -1252 -1517
27 : -1316 -1420 -1301 -1632 -1585 -1721 -1322
# Map BB_TESTMAP Measured Current
1 : -160 -148 -118 -130 -131 -150 -156
2 : -150 -125 -166 -138 -138 -162 -150
3 : -140 -146 -129 -148 -140 -139 -161
4 : -129 -155 -164 -161 -135 -157 -146
5 : -173 -140 -161 -164 -165 -130 -139
6 : -146 -130 -166 -140 -169 -162 -157
7 : -148 -122 -157 -136 -168 -147 -132
8 : -163 -149 -158 -142 -175 -146 -151
9 : -136 -165 -124 -142 -143 -170 -134
10 : -174 -130 -170 -126 -179 -152 -141
11 : -135 -135 -166 -127 -147 -158 -162
12 : -151 -164 -123 -170 -146 -163 -128
13 : -179 -138 -131 -145 -158 -159 -154
14 : -151 -137 -155 -141 -160 -168 -162
15 : -163 -172 -149 -149 -128 -164 -163
16 : -138 -167 -147 -169 -162 -171 -178
17 : -176 -161 -174 -171 -147 -147 -142
18 : -179 -143 -163 -129 -134 -132 -161
19 : -133 -162 -141 -172 -159 -140 -180
20 : -147 -167 -182 -114 -166 -186 -152
21 : -166 -196 -168 -166 -173 -144 -168
22 : -168 -155 -167 -179 -175 -145 -138
23 : -147 -150 -123 -164 -134 -145 -153
24 : -140 -171 -131 -167 -160 -183 -163
25 : -143 -127 -165 -156 -152 -164 -158
26 : -143 -158 -141 -155 -141 -132 -166
27 : -135 -152 -138 -173 -170 -184 -140
BigBite thresholds E dE(tdc) dE(trigger)
Wed May 7 16:11:49 EDT 2008 375 140 280
Threshold for 'mwdc_back' is +7.504 V (3072)
USAGE: ./threshold
-g|--get|get Get threshold for
-s|--set|set Set threshold for to
-v Verbose.
-h This help text.
is a voltage in units of mV or V
is from table below
name Channel Range Description
----- ------- ------- -----------
BB_Cer 7 -1V -- 0V BB Cerenkov Threshold
BB_PSh 6 -1V -- 0V BB Preshower Threshold
BB_Sh 5 -1V -- 0V BB Shower Threshold
mwdc_back 2 0V -- +15V Back MWDC Threshold
Using controller IP: bbcamac1.jlab.org. I/O reg in slot 12
Using cnaf command: './cnafb'
Examples:
./threshold -g mwdc_back
---> Return thresh for mwdc_back
./threshold -s BB_Sh -200
---> Set thresh for BB_Sh to -200 mV
./threshold -s mwdc_back 100mV
---> Set thresh for mwdc_back to 100 mV
./threshold -s mwdc_back 1.0V
---> Set thresh for mwdc_back to 1.0 V