We have changed the threshold in the Maximum number of hits per ADC channel in the crl and checked the increase of dead time:
RUN 4067 standard configuration: CRAMS_MAX_CHAN=120 DeadTime=9%
RUN 4068 CRAMS_MAX_CHAN=240 DeadTime=14%
RUN 4069 CRAMS_MAX_CHAN=180 DeadTime=12%
RUN 4070 CRAMS_MAX_CHAN=120 (BACK TO STANDARD) DeadTime=9%
keep going with threshold at 120