    Moller results on Feb 07 12:15-13:55

    E.Chudakov, A.Glamazdin

    Tuning took about 2h00m.


    Final results
     #    Wien angle  beam     runs target Ptarget% Pbeam% chi2
    1 30° 0.80 µA 13131-13138 5 8.03 -75.3±0.2±2.0 6.8
    2 30° 0.80 µA 13139-13142 4 7.50 -75.4±0.2±2.0 0.2
    3 30° 0.40 µA 13143-13146 2 8.23 -75.5±0.2±2.0 0.9
    average 1,2,3 30° average the targets -75.4±0.1±2.0 3.4

    The energy measured (HALLA:p) fluctuated, mostly during the 1-st measurement. The results with the target 5 more or less match this fluctuation. We are not sure how accurately the energy is measured at 0.8 µA. More studies are needed to extrapolate the polarization measured with no energy lock to the polarization during the regular runniing.