User name Kai Pan
Log entry time 23:04:03 on November 07, 2009
Entry number 298188
Followups:
keyword=PVDIS DAQ: delay timing, discriminator threshold, width
1. Now all 6 groups on left arm were added 113~116 ns delayed cables.
2. discriminator threshold:
Left: H:65,M:25,L:20
Right: H:90, M:20, L:20
3. discriminator width:
previously, the narrow on left arm is set to 60ns due to the deadtime
test before happex. I restore all narrow to be 30ns now. All wide were
100ns already. Right arm is 30ns in narrow and 100ns in wide already.
4. I found that on right arm shower sum8 of group2, there is one
unterminated output channel which caused many jitters when I was checking
the offset of TS analog sum. Problem fixed.
A copy of this log entry has been emailed to: xiaochao, rom, rsubedi, pkpan