A list of the relay-controlled devices for the target, along with the digital readbacks, is tabulated below. Refer to the ADIO-IP and VMIc223 manuals for pin-out designations.
| Device | Action | VMIc Relay | Digital Readback |
| Solenoid Valves | |||
| SV91108 | Open | 16 | ADIO 1, ZB0 |
| Close | 17 | ADIO 1, ZB1 | |
| SV91110 | Open | 18 | ADIO 1, ZC4 |
| Close | 19 | ADIO 1, ZB3 | |
| SV91111 | Open | 20 | ADIO 1, ZB4 |
| Close | 21 | ADIO 1, ZB5 | |
| SV91115 | Open | 22 | ADIO 1, ZB6 |
| Close | 23 | ADIO 1, ZB7 | |
| SV91115A |
Open | 23 | ADIO 1, ZB6 |
| Close | 22 | ADIO 1, ZB7 | |
| SV91148 | Open | 24 | ADIO 1, ZC0 |
| Close | 25 | ADIO 1, ZC1 | |
| SV91149 |
Open | 25 | ADIO 1, ZC2 |
| Close | 24 | ADIO 1, ZC3 | |
| OVC Controls | |||
| PV91104 | Open | 2 | ADIO 2, ZB0 |
| Close | 3 | ADIO 2, ZB1 | |
| OVP91130 | Start | 4 | ADIO 2, ZB2 |
| Stop | 5 | ADIO 2, ZB3 | |
| MP91132 |
Start | 6 | ADIO 2, ZB4 |
| Stop | 7 | ADIO 2, ZB5 | |
| Warm Return Valve | |||
| EV9177 |
Direction | 0 | na |
| Enable | 1 | na | |
The magnet exhaust/recycle valves, SV91115 and
SV91115A are controlled by a single 120 VAC relay. One
valve is controlled by the Normally Open set of contacts,
the other by the Normally Closed set. Therefore, only
one pair of 5 V VME relays and one pair of digital
inputs are necessary.
The refrigerator exhaust/recycle valves, SV91148 and
SV911149 are controlled by a single 120 VAC relay. One
valve is controlled by the Normally Open set of contacts,
the other by the Normally Closed set. Therefore, only
one pair of 5 V VME relays are necessary. Two
pairs of digital inputs are used.
MP91132, the backing pump for the oil diffusion pump, is not
connected to a relay, but is left running continuously. The
software controls do exist for remote control however.
EV9177, the warm return valve, is controlled by
two static (non-momentary) 5 V relays. The first sets a
direction (energize = CLOSE). The second enables valve
motion (energize = MOVE).