Privacy and Security Notice
elastic
Final H(e,e'p) results (?) 05/31/06
The HMS acceptance cuts are:
abs(hsyptar)<0.03.and.abs(hsxptar)<0.055.and.abs(hsdelta)<6.0,
highest Q2
abs(hsyptar)<0.04.and.abs(hsxptar)<0.07.and.abs(hsdelta)<8.0,
other two data points
The SOS acceptance cuts are:
abs(ssyptar)<0.07.and.abs(ssxptar)<0.04.and.abs(ssdelta-2.5)<12.5

Q2 (GeV2)
|
Y data (counts/mC)
|
Y simc (counts/mC)
|
Ydata/Ysimc
|
Stat. error
|
4.511
6.1
7.47
|
53.6289
11.5775
3.35994
|
51.323
10.9023
3.21941
|
1.04493
1.06193
1.04365
|
0.0101589
0.0167812
0.0136594
|
ssxptar 05/31/06
I tried increasing the ssxptar cut from
ssxptar<0.04 to ssxptar<0.045. John A. pointed out
that we should let the collimator define the acceptance. However,
the normalized yields increased by less than 1% with this change and
the yield ratios did not change to 5 decimal places.
I then tried reducing the ssxptar acceptance to ssxptar<0.020 to check the tails in
the hsxptar distributions.
Nominal cuts,
abs(ssxptar)<0.040

|
New cuts, abs(ssxptar)<0.020

|
It
appears that the tails in the hsxptar distribution move with the new
cut. Therefore, these tails are most likely due to resolution,
and should not be cut.
BCMs, Form factor parameterization,
etc. 05/31/06
I tried a few more things. I
removed the hsxptar cuts described below (05/31/06), and then used the
XEM BCM calibrations (the summer global fit for our July data and their
late November fit for our December data).
Q2 (GeV2)
|
Ydata/Ysimc before
|
Stat. error
|
Ydata/Ysimc after
|
Stat. error |
4.511
6.1
7.47
|
1.04506
1.06413
1.0766
|
0.0101609
0.0168231
0.0140541
|
1.04493
1.0652
1.06693
|
0.0101589
0.0168329
0.0139641
|
I
then changed the form factor parameterization (John A.'s Rosenbluth
form factors), because I noticed there was a mistake. Above Q2=6
GeV2, the G_E elastic form factor needs to be set to G_dipole (or what
appears to be very close to it). This form factor does not
contribute significantly at large Q2, however, the parameterization may
become very large in this region.
Results
with the corrected form factor parameterization
Q2 (GeV2)
|
Ydata/Ysimc
|
Stat. error
|
4.511
6.1
7.47
|
1.04493
1.06193
1.04365
|
0.0101589
0.0167812
0.0136594
|
I also tried Peter Bosted's
parameterization again. However, the yield ratios all moved up by
<1%.
Here is the plot of the ratio after changing back to John A.'s
parameterization. Note- the nominal hsxptar acceptance was used
here, and in the previous two tables.

Hsxptar cuts 05/31/06
I tried using hsxptar cuts to remove the
long tails in the experimental data hsxptar distributions.
Previously, I had hsed hsxptar cuts to remove the edges of both the
SIMC and data distributions.
This is how hsxptar looks with the new cuts :
abs(hsxptar+0.0025)<0.0225, at the middle and low Q2
abs(hsxptar+0.00125)<0.0175, at the highest Q2

The yield ratios with the new cuts are

Hsxptar cuts 05/30/06
The hsxptar cut abs(hsxptar)<0.15
seems to make the data yield agree better with SIMC.
I didn't really get anywhere with this, but this is what I have so far.
1)
The coplanarity distribution is broader in the data compared to
simc. I am not sure why -see the work below (05/29/06).
2) Both cuts hsxptar<0.15 and hsxptar>-0.15 increase Ydata/Ysimc,
however, the former is worse.
Nominal
hsxptar cuts
Q2
Ydata/Ysimc Stat error
4.511 1.04506
0.0101609
6.1 1.06413
0.0168231
7.47 1.0766
0.0152574
hsxptar<0.015 ... This
has the worst problems
Q2
Ydata/Ysimc Stat error
4.511 1.02616
0.0101572
6.1 1.04258
0.0166829
7.47 1.06588
0.01399
hsxptar>-0.015
Q2
Ydata/Ysimc Stat error
4.511 1.01099
0.0106755
6.1 1.05829
0.0173991
7.47 1.02881
0.0138151
Tight cut on hsxptar
(abs(hsxptar)<0.15)
Q2
Ydata/Ysimc Stat error
4.511 0.988393
0.0106613
6.1 1.03519
0.0172414
7.47 1.01777
0.013746
3) The events with hsxptar<-0.15 seem
to be reconstructed correctly
Systematic error 05/30/06
The significant contributions are:
Proton absorption: 0.5% random and 2.0% scale
HMS fid effic: 1% random and 1.0% scale
SOS fid effic: 0.5% random and 0.5% scale
Radiative corrections: 0.5% random and 1.0% scale
Acceptance:
2.0% scale
Total systematic uncertainty: 1.3% random and 3.2% scale
Model dependence: 2-3% at the lowest Q2, and 3-4% at the other two
points.
In the plot below, the solid line represents the random and scale
uncertainty added in quadrature, while the dashed line represents the
random systematic uncertainty. The model uncertainty has not been
included. The error bars on the data points are the statistical
uncertainty.

When
an extra cut is included to remove the edges of the hsxptar
distributions in both the data and SIMC, abs(hsxptar)<0.015, the
ratios become

Coplanarity 05/29/06
In an attempt to find out why the hsxptar
distributions have longer tails in the data compared to SIMC (see
"Rebinned hsxptar plots 05/29/06" below), I plotted
lab
phi of the particle in the SOS - lab phi of the particle in the HMS -
180 deg
which should be zero.
The Lab phi angle was calculated from
myhsphi=atan2(hsxptar, (sin(hs_spec_angle)-hsyptar*cos(hs_spec_angle))
)
myssphi=atan2(ssxptar, (-sin(ss_spec_angle)-ssyptar*cos(ss_spec_angle))
)
if
(myssphi<-1.5) then
myssphi=myssphi+2.0*3.14159265358
endif
where, hsxptar has, in addition, the
appropriate oop offsets.
In any case, these plots of the coplanarity show that the width of the
SIMC distribution is smaller than the data by about 0.5 degrees, which
corresponds to 9 mrad. This discrepancy could be due to (1)
resolution (2) multiple scattering (3) peaking approx.

Proton absorption 05/29/06
Ok, after correcting for the proton
absorption once only ...
Q2 (GeV2)
|
Ydata/Ysimc
|
Stat. error
|
0.633
4.511
6.1
7.47
|
0.781453
0.988393
1.03519
1.0184
|
0.00396284
0.0106613
0.0172414
0.0137499
|
However, the results above use the
nominal acceptance cuts, with a tight cut on hsxptar
abs(hsxptar)<0.015
When I use the nominal acceptance cuts (without the tight hsxptar cut),
abs(hsyptar)<0.04.and.abs(hsxptar)<0.07.and.abs(hsdelta)<8.0
abs(ssyptar)<0.07.and.abs(ssxptar)<0.04.and.abs(ssdelta-2.5)<12.5
I obtain the following yield ratios
Q2 (GeV2)
|
Ydata/Ysimc
|
Stat. error
|
0.633
4.511
6.1
7.47
|
0.817759
1.04506
1.06413
1.07826
|
0.00364188
0.0101609
0.0168231
0.0140651
|
HMS Cerenkov cut 05/29/06
I tried using a hcer cut to remove pions,
without any corrections for the good protons removed due to knock-on
electrons and Cerenkov blocking. The cut was hcer_npe<0.3.
Q2 (GeV2)
|
Ydata/Ysimc
|
Stat. error
|
4.511
6.1
7.47
|
1.0177
1.06676
1.04622
|
0.0111092
0.0179923
0.0143205
|
However, e-pi coincidences should not be
present due to the cointime cut, and the W cut, abs(W-0.94)<.1.
Furthermore, knock-on electrons have been ignored. Based on the
rejection ratio measured for pions in the HMS Cerenkov for the cut
hcer_npe>0.5, which was around 50:1, the expected fraction of
protons that produce knock-on electrons is about 2%. This is
about the size of the effect seen above.
Jason's write-up on knock-on electrons is here
http://hallcweb.jlab.org/xmptlog/0409_archive/040902160325.html
Hee'p offsets 05/29/06
The plots for the previously fitted heep
offset results (see 05_28_06 below) are shown below. SIMC with
radiation, eloss, etc turned on is shown in blue. The data are
the green and red histograms, which should overlap.
Q2=0.633 GeV2
Q2=4.51 GeV2
Q2=6.1 GeV2
Q2=7.47 GeV2
The offsets were:
Ebeam(GeV)
|
P hms(GeV)
|
TH hms(deg)
|
P sos(GeV)
|
TH sos(deg)
|
dE
|
dth_ss
|
dp_ss
|
dth_hs
|
dp_hs
|
dssxptar
|
dhsxptar
|
2.038
|
-1.692
|
25.005
|
0.874
|
54.90
|
0.59
|
0.0
|
0.48
|
0.6
|
0.9
|
-0.0032 |
-0.0011
|
5.012
|
4.116
|
18.70
|
-1.7227
|
50.00
|
0.79
|
-1.0
|
11.0
|
0.6
|
0.9
|
-0.0032 |
-0.0011 |
5.766
|
4.867
|
16.19
|
-1.6868
|
51.61
|
0.2
|
0.0
|
8.11
|
0.0
|
0.9
|
-0.0032
|
-0.0011 |
4.021
|
3.232
|
22.15
|
-1.588
|
50.00
|
-0.75
|
-1.0 |
3.85
|
1.0
|
0.9
|
-0.0032
|
-0.0011 |
The
format of the offsets "d[variablename]" are the same as those in
"heepcheck.f". For example, for the electron in the sos
p_e = ssp*(1-dp_ss/1000.0)
th_e =
mysstheta-dth_ss/1000.0 [rad]
p_p =
hsp*(1-dp_hs/1000.0)
th_p =
myhstheta-dth_hs/1000.0 [rad]
newE=e_beam*(1+dE/1000.0)
newssxptar= ssxptar- dssxptar
[rad]
newhsxptar=hsxptar-dhsxptar [rad]
Acceptance edges 05/29/06
Previously (see 05_24_06 below), I tried
using cuts to select a narrow region in the center of the
acceptance. Now, the effect of removing events near the edges of
the acceptance was measured.
Q2 (GeV2)
|
Ydata/Ysimc before
|
Stat. error
|
Ydata/Ysimc after
|
Stat. error |
0.633
4.511
6.1
7.47
|
0.826934
1.04592
1.09544
1.07767
|
0.00419348
0.0112818
0.0182449
0.0145502
|
0.793018
1.05639
1.07809
1.08236
|
0.00424169
0.0118871
0.0193476
0.0158556
|
The cuts used were:
abs(hsyptar)<0.03.and.abs(hsxptar)<0.015.and.abs(hsdelta)<6.0
abs(ssyptar)<0.055.and.abs(ssxptar)<0.03.and.abs(ssdelta-2.5)<10.
Rebinned
hsxptar plots 05/29/06

The hsxptar distributions in the last
three Q2 settings appear rounded, but the shape is not too different
from the SIMC (blue) distributions.
Calculation of the data norm. yields
05/29/06
The links below show the values of the
the efficiencies, etc. used in the calculation of the normalized
yield. An example is provided below.
list1
list2
list3
list4
setting names
el1lh2, Q2 = 0.633 GeV2
el2lh2, Q2 = 4.511 GeV2
el3lh2, Q2 = 6.1 GeV2
el4lh2, Q2 = 7.47 GeV2
Example, run 49537 (Q2=6.1 GeV2, electron in the SOS):
yield = synccorr*blockcorr/scereff*cpre/ctrg*hslowfid*scerfid
/(1-helecdt/100)/(1-selecdt/100)/Tproton*(pcounts-bcounts)
=
1.0*1.0/0.997471*1356/1356*/0.9300/0.9929
/(1-0.0001)/(1.0)*1.058*(627-0)
= 720.291
Which can be seen to be equal to the
value determined by the analysis code (720.298).
Definitions:
scereff = efficiency of the sos cerenkov
hslowfid = fid efficiency of the HMS for slow particles
(protons). The cuts used were (1) small or zero hcer signal (2)
calorimeter signal above a low threshold (3) event type was a
coincidence.
scerfid = fid efficiency of the SOS with a cut to select large scer and
calorimeter signals
helecdt = electronic dead time (%)
1/Tproton = 1/(proton transmission), the correction for proton
absorption
The normalized yield for this setting is determined from the corrected
bcm2 (ie. cbcm2). Charge cuts were used for run 49537,because
there was a trip in this run. Most runs do not have trips, and so
cbcm2 = bcm2.
norm yield = 720.291/ (72000/1000) = 10.004 counts/mC
Plots showing the normalized
yield for each run.
The setting averaged normalized yields (with dummy subtraction) are:
Q2 (GeV2)
|
xnyield
|
0.633
4.511
6.1
7.47
|
73578.
43.1746
10.3873
3.14522
|
More checks 05/28/06
The results did not change significantly
when the sync filter was used
Q2 (GeV2)
|
Ydata/Ysimc
|
Stat. error
|
0.633
4.511
6.1
7.47
|
0.828136
1.04829
1.10506
1.05668
|
0.00419797
0.0112964
0.0183294
0.0143972
|
I
then played with the heep offsets a little more to get the peaks
centered even closer to SIMC.
Q2 (GeV2)
|
Ydata/Ysimc
|
Stat. error
|
0.633
4.511
6.1
7.47
|
0.826607
1.04597
1.09549
1.07767
|
0.00419182
0.0112823
0.0182457
0.0145502
|
Here are some plots of ssshtrk (electrons
in the SOS) and hcer_npe (protons in the HMS), and this is why these
cuts on these variables have not been applied in the analysis.
The background from misidentified ep elastic events appears to be on
the % level, and it seems that using cuts on these variables will
unecessarily complicate the analysis (one needs to coincider Cerenkov
blocking and knock-on electrons, for example).
Q2= 4.511 GeV2
Q2=
6.1 GeV2
Q2=
7.47 GeV2
More checks 05/27/06
The analysis was redone using Tanja's
offsets (Table 3.2, page 107 of her thesis). dphms=-0.13% and
dtheta = 0 in both the hms and sos.
Q2 (GeV2)
|
Ydata/Ysimc
|
Stat. error
|
0.633
4.511
6.1
7.47
|
0.884794
1.1
1.12457
1.10806
|
0.00392917
0.0107005
0.017709
0.0146358
|
Then, I refit the offsets beginning with
Dave's offsets (dphms=-0.09% and dthetahms=-0.6mrad), and made
some small adjustments to get the peaks to line up with simc.
Q2 (GeV2)
|
Ydata/Ysimc
|
Stat. error
|
0.633
4.511
6.1
7.47
|
0.87798
1.10472
1.13634
1.10915
|
0.00389827
0.0107451
0.0178939
0.0146505
|
Based on the entry below on 05/26/05, a
tight hsxptar cut was tried to remove the long tails seen in the
data. The highest Q2 has the thinnest hsxptar distribution so the
cut was:
abs(hsxptar)<0.15
Q2 (GeV2)
|
Ydata/Ysimc
|
Stat. error
|
0.633
4.511
6.1
7.47
|
0.839753
1.0484
1.10683
1.056
|
0.00424456
0.0112971
0.0183449
0.0148575
|
More checks 05/26/06
Latest hsxptar distributions (first three
are the delta scan, last three are the smallest Q2 to largest Q2
settings). Blue = simc.

Check of the random coincidences. There are no random
coincidences.
Large Q2

Small Q2

Cuts
used in the hee'p analysis 05/26/05
Data cuts
PID and missing mass cuts
(scer_npe>1.35.and.Em<0.025.and.abs(W-0.94)<0.1)
Acceptance
abs(hsyptar)<0.04.and.abs(hsxptar)<0.07.and.abs(hsdelta)<8.0
(abs(ssyptar)<0.07.and.abs(ssxptar)<0.04.and.abs(ssdelta-2.5)<12.5)
Jochen's
cut (Pg 80 of his thesis)
(ssyptar>-125.0+4.25*ssdelta+64.0*ssytar-1.7*ssdelta*ssytar)
(ssyptar<125.0-4.25*ssdelta+64.0*ssytar-1.7*ssdelta*ssytar)
Coincidence blocking and sync cuts (values from the database)
hsrawct>[myblockcut].and.(hsrawct+ssrawct)>[mysynccut]
Angle cointime cuts (protons in the SOS)
abs((cointime)-[a])<[b].and.(ssbeta)>[b1min].and.(ssbeta)<[b1max]
(ssbeta)>[b2min].and.(ssbeta)<[b2max]
(cointime)-[a]>-[b]+[m]*(-0.5)*((ssbeta)-[c]-abs((ssbeta)-[c]))
(cointime)-[a]<[b]+[m]*(-0.5)*((ssbeta)-[c]-abs((ssbeta)-[c]))
SIMC cuts
Em<0.025.and.abs(W-0.94)<0.1
abs(hsyptar)<0.04.and.abs(hsxptar)<0.07.and.abs(hsdelta)<8.0
abs(ssyptar)<0.07.and.abs(ssxptar)<0.04.and.abs(ssdelta-2.5)<12.5
(ssyptar>-125.0+4.25*ssdelta+64.0*ssytar-1.7*ssdelta*ssytar)
(ssyptar<125.0-4.25*ssdelta+64.0*ssytar-1.7*ssdelta*ssytar)
The latest Y_data/Y_SIMC 05/25/06
Q2 (GeV2)
|
Ydata/Ysimc
|
Stat. error
|
0.633
4.511
6.1
7.47
|
0.87454
1.11799
1.15437
1.17445
|
0.00388384
0.0107887
0.0179889
0.0152029
|
Checks
of elastic analysis 05/24/06
I checked the data normalized yields using the paw interface
(independent of my existing kumacs). The data yields I have been
using are in reasonable agreement with the simplified
analysis. One place to start looking may be in the HMS fid effic,
which was 0.93 at some settings. If this is artificially small,
it may partially explain the higher data normalized yields.
Otherwise, I can not see any 10% discrepancies in the analysis of the
experimental data. I used the hadron fid (87-95%) for the proton
arm and the electron fid (99.3-99.5%) for the electron arm.
I
replayed the elastic runs with slightly different fid track effic
calculations. I required that there is a small (non-zero) signal
in
the hadron-arm calorimeter. The fid effic was essentially
unchanged.
As before, I still use the coincidence hadron fid efficiency (Cerenkov
< 0.2 npe) for the hadron arm.
The W and xbj plots between data and SIMC have reasonable
agreement. This suggests that the PID cuts are reasonable.
W plots for Q2=0.633-0.64 and
for Q2=4.511-7.47
xbj
plots for Q2=0.633-0.64 and
for Q2=4.511-7.47
Next, I checked that the infiles kinematics were set correctly.
This could be done by plotting hsp (data vs SIMC), ssp, hstheta and
sstheta. If the central kinematics were set incorrectly, these
histograms will not match. The results below show that the
central kinematics were, in fact, set correctly. The plots below
are normalized so that SIMC has the same area as the data.
de4lh2, Q2 = 0.633
de5lh2, Q2 = 0.637
de6lh2, Q2 = 0.64
el2lh2, Q2 = 4.511
el3lh2, Q2 = 6.1
el4lh2, Q2 = 7.47
Increasing the generation volume in SIMC (more events outside of the
acceptance) did not seem to change the results:
Q2 (GeV2)
|
Ydata/Ysimc before
|
Stat. error
|
Ydata/Ysimc after
|
Stat. error |
0.633
0.637
0.64
4.511
6.1
7.47
|
0.835908
0.959927
0.87903
1.06541
1.09743
1.15238
|
0.00294076
0.00404626
0.00800635
0.00910868
0.0148896
0.012746
|
0.838134
0.962656
0.870977
1.06562
1.1014
1.14734
|
0.0029529
0.00406916
0.00794429
0.00911016
0.0149438
0.0126902
|
Using an Em cut of EM<25MeV (applied to both the data and simc)
helped a little, expecially with the highest Q2 point, but there are
still problems. The delta scan runs are now worse.
Q2 (GeV2)
|
Ydata/Ysimc before
|
Stat. error
|
Ydata/Ysimc after
|
Stat. error |
0.633
0.637
0.64
4.511
6.1
7.47
|
0.835908
0.959927
0.87903
1.06541
1.09743
1.15238
|
0.00294076
0.00404626
0.00800635
0.00910868
0.0148896
0.012746
|
0.809157
0.903222
0.784015
1.05697
1.09139
1.11027
|
0.00306525
0.00406864
0.00750922
0.0101257
0.0168878
0.0142297
|
I
finally got somewhere! I cut on only the central part of the
acceptance, and the yield ratio came out very close to one. The
only runs with problems are the delta scan runs, that are outside of
the central part of the acceptance.
Q2 (GeV2)
|
Ydata/Ysimc before
|
Stat. error |
Ydata/Ysimc before |
Stat. error |
4.511
6.1
7.47
|
1.05697
1.09139
1.11027
|
0.0101257
0.0168878
0.0142297
|
1.00957
1.02392
1.03172
|
0.0225653
0.0354379
0.0281909
|
However, it seems that this was just a
statistical fluctuation.
I found a problem with the way I was implementing the proton
absorption. Previously, I applied to absorption correction
(1/0.945) to the data whenever histograms were produced, and not when
the yield ratios were calculated. As the SIMC histograms were
normalized to match the data, the absorption correction was having no
effect. After implementing the correction, the yield ratios are
now worse.
Q2 (GeV2)
|
Ydata/Ysimc
|
Stat. error
|
0.633
4.511
6.1
7.47
|
0.87454
1.11799
1.15437
1.17445
|
0.00388384
0.0107887
0.0179889
0.0152029
|
Elastic
H(e,e'p) yield check 05/23/06
SIMC was run again, with John A's parameterization, and plots of
spectrometer quantities were produced. All SIMC plots are
normalized to match the data. There was almost no change in the
ratio of the yields with the new form factor parameterization.
de4lh2, Q2 = 0.633
de5lh2, Q2 = 0.637
de6lh2, Q2 = 0.64
el2lh2, Q2 = 4.511
el3lh2, Q2 = 6.1
el4lh2, Q2 = 7.47
More plots vs. spectrometer variables are here
de4lh2, Q2 = 0.633
de5lh2, Q2 = 0.637
de6lh2, Q2 = 0.64
el2lh2, Q2 = 4.511
el3lh2, Q2 = 6.1
el4lh2, Q2 = 7.47
Elastic H(e,e'p) yield check 05/23/06
Elastic runs were analyzed and simulated with SIMC.
The data normalized yields were divided by the proton transmission
(0.945 in the HMS and 0.95 in the SOS).
Peter Bosted's parameterization of the form factors was used in these
results (and changed to John's parameterization in subsequent analysis).
The settings are (table of settings)
*
Delta scan for elastic e,e'p, Q2 = 0.646 GeV2, beam=2038, Electron in HMS
de4lh2 48865 48866
de5lh2 48870
de6lh2 48872
* electron in SOS, Q2 = 4.56 GeV2,
beam=4021
el2lh2 49874 49875 49876
49877
* electron in SOS, Q2 = 6.17 GeV2,
beam=5012
el3lh2 49532 49533 49534
49535 49536 49537
* electron in SOS, Q2 = 7.37 GeV2,
beam=5766
el4lh2 52394 52395 52396
52397 52398 52399 52401 52402 52403 52411 52412 52413 52414 52415 52416
The first results for the yields are shown below
Q2
Ydata/Ysimc Stat.Error
0.633 0.842488 0.0029639
0.637 0.967865
0.00407972
0.64
0.886777 0.00807693
4.511
1.07136
0.00915955
6.1
1.1012
0.0149407
7.47
1.1361
0.0125656
Plots of Q2 for the settings above, in this order, are shown
below. Red=data, Blue=simc. The agreement between data and
SIMC is poor, and the error bars for the data in the first three plots
appear to be too small compared to the point-to-point variation.
The SIMC plots have been normalized to match the data.

