Ruthenium Oxide:
Each stick is equipped with two RuOsensors located
above the deuteron and proton cells.
For T < 6 K the temperature-resistance relation is the following
T = (A /(R - B))**2
where R is the measured resistance in Ohms and A, B are given in the following
table.
Typically at liquid helium R>1500 Ohms.
| STICK NUMBER | POSITION | A (Ohms*K**1/2) | B (Ohms) |
| 1 | top | 833 | 860 |
| 1 | bottom | 861 | 850 |
| 2 | top | 947 | 872 |
| 2 | bottom | 883 | 874 |

Cernox 1050:
Each stick is equipped with a cernox located above the 4 target cells and
used as reference temperature for
the two thermocouples.
The calibration of the cernox for both sticks is stored in the ITC 503
temperature controller, channel
number 2.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Thermocouples:
each thermocouple (two for each stick) consists of two junctions (Chromel/Au/Chromel):
the reference
junction is located near the top cernox
(1050), while the sensing is placed inside the target container.
The measured voltage is linearly related to the temperature difference
of the two junction. The calibration
given below was obtained when the reference junction
is mantained in liquid nitrogen ( T = 77K). In this
condition the thermoelectric constant s (microvolt/kelvin)
is:
| STICK NUMBER | POSITION | S |
| 1 | top | 18.86 |
| 1 | bottom | 18.86 |
| 2 | top | 21.60 |
| 2 | bottom | 21.73 |

Cernox 1030:
Each stick is equipped with a cernox located at the bottom and used to
monitor the helium level and temperature.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|


CONNECTIONS OF THE STICK SENSORS
All the temperature sensors are wired onto a 15 pins
connector located in the inner part of the stick top flange. The connection
to the outside is provided by a 24 pins Fisher connector. The following
tables summarize the present wiring.
STICK #1 (STICK #2)
| SENSOR | RESISTANCE AT ROOM TEMPERATURE (Ohms) | 15 PIN CONNECTOR | 24 PIN CONNECTOR | REDOUT INSTRUMENT |
| RuO top | 1016 (1014) | 1 | 20 | OXFORD resistance bridge |
| RuO top | 2 | 19 | channel 5 | |
| RuO bottom | 1016 (1029) | 3 | 18 | OXFORD resistance bridge |
| RuO bottom | 4 | 17 | channel 6 | |
| Cernox I+ | 9 | 11 | ITC 503 ch 2 | |
| Cernox V+ | 57 (81.5) | 10 | 12 | |
| Cernox V- | 8 | 14 | ||
| Cernox I- | 7 | 13 | ||
| Thermocouple top | 67 (31) | 13 | 4 | to be defined |
| Thermocouple top | 14 | 3 | ||
| Thermocouple bottom | 131 (31) | 11 | 6 | to be defined |
| Thermocouple bottom | 12 | 5 | ||
| Heater top | 25 (46) | 5 | 24 | ITC 503 ch 2 |
| Heater top | 15 | 23 | ||
| Heater bottom | 25 (24) | 6 | 22 | ITC 503 ch 2 |
| Heater bottom | 15 | 23 |
(plug view)
View of the 24 pin connector on the stick upper flange