Privacy and Security Notice
epics variable in baryon
Privacy and Security Notice
Privacy
and Security Notice
Epics variables
during Baryon experiment
General
Info
Epics information from various equipment is stored with
the data. What data is stored is determined by the file epics.dictionary
during data taking. In the beginning of the file are comments
explaining the format. Most epics variables have comments explaining
what they are. Some variables are saved every 2 seconds and some every
30 seconds.
When replaying the epics info can be written out to a
file, epics####.txt. For most runs this was done during the
online replay and the epics files are in
/work/hallc/e01002/charge_scalers. For subsequent replay the epics
files are in /work/hallc/e01002/pass#/scalers .
A
tcl script , extract_epics.tcl
,which goes through the epics files and extracts the epics
info that one wants is available. The script extracts the data
for each "epics read" and then get the mean and standard deviation for
the epics variable for the run. The info is stored in the file
epics.out .
SOS
magnets
Plots of the SOS Q1,
D1 and D2 magnet currents versus run
number in sos_magnets_p=1.0.ps
and sos_magnets_p=1.74.ps
. For the 1.74 setting, the currents are unstable for Q1
and D1 at the beginning of
the experiment but after that they are stable. For pass1
and pass2 replay, all runs which had SOS nominally set to 1.74 used
spcentral = 1.74 in the standard.kinematics file. For pass 3 replay,
the spcentral was modified for each run by spcentral = 1.74*(D1
current)/985.5 . A plot
of the ratio (D1
current)/985.5 versus run number. The values of spcentral used are
given in this file
( listing: run number, spcentral, standard
deviation of spcentral).
Beam Energy
The
beam energy was not measured during the BARYON experiment. A table of
the beam energies during the Spring03 running is here.
The beam energy should have been
5.5015 GeV as it was during the
measurement on April 24th, 2003. There was a brief change in the linac
energy between this measurement and the start of the BARYON experiment,
but then the linac energy was suppose to be set to the previous
value. One can see in the table of beam energies that there is a
epics variable HALLC:p which is a measurement of the beam energy
determined by accelerator ( using ARC dipole setting and beam
positions) . The
HALLC:p value is usually very close to the measured value using
the ARC
beam energy technique. A plot
of
the value HALLC:p versus run number shows
that at the start the beam energy is dropping. After around run 46260
the beam energy is fairly stable at around 5499 GeV. The "error" in the
plot is the standard deviation of all the measurements of
HALLC:p during the run. For some run periods the standard deviation is
almost zero which probably indicates that the epics values of
HALLC:p wasn't being updated. There is a file of
the beam energy values ( run number, HALLA:p, std dev. , HALLC:p, std
dev. ) .
During
pass 3 replay a value of 5496 GeV was used for the beam energy for all
runs.
Beam position.
Plots of the beam
positions at BPMA
(z = 327.15cm) , BPMB
(z
=231.46 cm)
and BPMC
(z =
135.35cm)
. During the experiment, we tried to keep position at BPMC
constant. It was fairly stable except for a couple of regions. Y is the
vertical beam position with +Y up. X is the horizontal beam
position with +X towards beam right.
Projecting
to the target ( z= 0 ) to get the target positions one can use
different combinations of the BPMs. A plots of the target position
using BPM AC
, AB
, and BC
combinations have been made. A plot
comparing are three combinations shows that
different combinations are not consistent with each other and that
offsets in the relative positions of the BPMs are needed to get
them to give the same target positions. But the different combinations
show the same relative behavior. The X ( horizontal position) is fairly
constant ( except for a dip at runs 46260-46290), but the Y
(vertical position) has a couple of periods of different
positions.
Run periods
|
Vertical
position
|
46068-46259
|
0.15mm
|
46260-46293
|
0.40mm
|
46294-46319
|
0.10mm
|
46320-46804
|
0.45mm
|
The vertical position effects
delta reconstruction. It is 0.077 % delta per mm offset in vertical
position. So if we don't put in the vertical beam position then
if the beam is off by 1mm then delta would shift by .077% ( or almost
10-3) [ for SOS this is .039%/mm ]. We corrected for the raster but we
didn't include movements in the centroid of the beam during replay ( up
through pass 3) .
Plots of the beam angle using BPM AC
, AB
, and BC
combinations have been made.
A plot
comparing are three combinations shows similar
discrepancies between the different combinations as seen for the
positions.