Goal
To measure the ratio of proton's electric to magnetic form factor to
2%
-
Beam Energy
1.67 GeV
-
Carbon Blocks
3"
-
Beam position
less than 0.25 for x and y for 3A and 3B
-
Raster size
2.0x2.0mm in SPOT .
-
Beam Current
2 microA. or maximum possible with deadtime < 30%. Expect rate of
2,200 hz at 2 microA.
-
Target
15cm LH2
-
Spectrometer settings
| Spectrometer |
Momentum |
Angle |
Collimator |
| Electron |
1.438 GeV |
24.45° |
open or 6msr |
|
|
|
|
| Proton |
0.6946 GeV |
58.97° |
open or 6msr |
-
Data taking
-
The collimator setting will
be decided by doing quick runs with the open collimator and the 6msr.
If the raw T5 rate increases
by more than 20% with open collimator then use open. Otherwise use 6msr
setting.
-
The T5 prescale should be
one.
-
The prescales for T1,T2,T3
and T4 should be set to give less than 10 Hz trigger rate for each. The
T8 prescale should be set to maximum.
-
The trigger setup needs
to be set for proton momentum.
-
Need to modify the header
file : hdr_fpp_89028_q04h
-
The data taking time
is estimated to be 3.0 hours. Will need to update after looking at carbon
scattering efficiency.
-
Need 2.2x10+6
total triggers which had scattering between 5 and 20 degrees
in carbon .