Privacy and Security Notice
Magnets Group Manual
TJNAF Experiment 89-033
Magnets Group Manual for the Experiment E89033
Table of Contents
General Overview
The HRS is composed of three superconducting
quadrupole magnets Q1,Q2 and Q3 and one superconducting dipole magnet.
A particle first traverses Q1 and Q2, then the dipole magnet and finally
Q3.
Cryogenic information about each magnet as well as control interface
for the power supplies is available on the control screen in Hall A counting
house. You can get there from the startup display
(described below). The power supplies themselves are located on the
gantry adjacent to the magnets.
To run the control interface for magnets,
- Logon on the hac computer (account hacuser, password
*****) and type HAC_hp if you are on HP machine or
HAC_xt if you are on X terminal.
- Startup display will show up:
It summaries information about all the
magnets. To get information on magnet power supplies, gaussmeters and NMRs
click the corresponding buttons. From this screen you can also get some
detector information screens.
To start an alarm handler for magnets type alh in hacuser
home directory (that is where you start).
General Task List
The main task of the magnet operator is to be sure that the
magnetic fields
and currents in the spectrometers are correct at all times!
- Fill out run sheet and print magnet summary screen at the beginning
of each run.
- Watch out for magnet power trips and communication software failures
and try to fix it if it happens (see here). Call
run corrdinator first if
you have to get an access to the hall to reset a power supply.
- Change magnet settings when it is required. There will be two of them
during the run time: first, for hydrogen kinematics and then, for oxygen
kinematics. See here how to do it.
- Watch out for helium level and lead flow and call a
Hall A technical support on-call if it would go low. See
here for helium
levels and lead flows limits for each magnet.
- Bring all the magnets down in case of a long interruption in data
taking (e.g., during the maintanance days and/or some other major
shutdowns).
Field Setting Instructions
Dipole Field Setting
To power up magnets:
- Go to the "Dipole Sytem" window (
electron or hadron);
- Push ON button and wait untill Power Supply will be ON
and Magnet will be Engaged (it will take 2-3 minutes);
To set desired field do the following:
For the Electron Dipole:
- Go to the "Electron Arm Dipole Sytem"
window;
- Type in desired value of current in Current Set
(yellow field) and hit Return;
- Go to the
"HRS Electron Dipole NMR" window;
 |
- Set the probe, make sure it has a right polarity;
- Set compensating coils voltage appropriately(if it was not set before),
see writeup by J.LeRose;
- Put NMR in Manual mode;
- Type in DAC number for the field being measured, see writeup by
J.LeRose
;
- Put NMR in {\bf AUTO} mode;
- Wait for desired current in Iout (about 20-30 minutes), in
"Dipole System" window, and then for a lock
in the NMR window("L" on the yellow background). If something
goes wrong, go here.
- When you got lock in
"HRS Electron Dipole NMR window, verify it with ocsilloscope
(in the middle room)
- Do the same procedure for the next probe;
|
- To find the field set look at the
"HRS Electron Dipole NMR"screen or
summary screen.
Remeber: Field in sec.5 "Magnet Settings"
is an average of two probes, high and low.
For the Hadron Dipole:
- Go to the
"HRS Hadron Dipole NMR" window;
 |
- Press Reg.Enable:YES;
- Press Search:ON;
- Type desired field into the Field Set and hit Return;
- Wait 20-30 minutes.
- When you got lock in
"HRS Hadron Dipole NMR window, verify it with ocsilloscope
(in the middle room)
|
To shut either of the dipoles down:
Go to the "Dipole Sytem" window (
electron or hadron) and push
"0 Amps" button. It will take 20-30 minutes to get there.
Quads Field Setting
To power up magnets:
- Go to the "Magnet Sytems" window (
Q1,Q2,Q3
for the electron arm and Q1,
Q2,Q3
for the hadron arm);
- Push Supply:ON button;
- Set polarity;
To set the desired current (in a corresponding "Magnet Systems"
window):
- Type in an absolute value of desired current in Current Set;
- Wait for desired current for about 10 - 15 minutes). If something goes
wrong, see sec.4.
- To find the field set look at the summary
screen or corresponding gaussmeter
screen.
To shut either of the quads down:
Go to the "Magnet System" window and push
the Supply:OFF. It will take 10-15 minuts to get there.
Possible problems and Their Solutions
NOTE 1: You can find the complete list of
all the possible
problems and how to react on them either in J.LeRose writeup or in general
OPS manual, both are in the counting room. Here is just a short list of
most probable problems.
NOTE 2: Occasionally, if you run into
communication problem (hardware
fails to communicate with software, some examples described below) you may
want to reboot VME crate. However, please, use it only if everything else
fails.
NOTE 3: If you are in doubts about what to
do, call an expert or Hall A tech support on-call.
How to Reboot:
-
- Prefered! Push one of the green buttons on the distribution box
in the middle room: for electron arm - bottom raw, left column,
for hadron arm - bottom raw, right column;
OR
- Rlogin on hallasc16(for Hadron Arm) or on hallasc14
(for Electron Arm) on account hacuser and password *******;
- Push Cntrl-X: all windows will go blank
- Make sure field regulation is enabled and set the dipole field as
described above;
- Set the Q2/Q3;
Q1 Problems
- Output current is equal to zero but field seems to be ok:
communication failure, power supply is ok:
- Look at the camera screen to make sure that you really do not have
any faults. If you do, see below;
- Push Reset button in a magnet control
screen. Wait untill it
complete reset procedure (red rectangular near the button will dissapear).
Try it couple times if it does not work first.
- If everything fails, reboot.
Recommendation for data taking:
Unless you rebooted, no need to interrupt data taking. If you rebooted,
stop the run beforehand.
- Output current and field are equal to
zero: magnet tripped:
- Look at the camera screen and make sure that this fault is from DL
section (slow discharge). If it is in DR section you must call a tech support.
- Push Reset button in a magnet control
screen. Wait untill it
complete reset procedure (red rectangular near the button will dissapear).
- Re-enter current setting, if it is required.
- If everything fails, call a Hall A tech support on-call.
Recommendation for data taking:
Stop data taking as soon as you saw the magnet tripped.
- If magnet starts to trip too often because lead flow is low,
you might want to call cryogenics and ask them to increase a pressure.
Q2/Q3 Problems
Dipoles Problems
- NMR is not locked and current is going in the wrong direction:
software confused:
Push Search:ON in the "HRS Dipole
NMR" window. NMR should lock in a while. For detailed instructions, see
here;
- On the camera display M3/M4 temerature is red: make a note
in a logbook and continue to take data;
- Field and current are zeros: magnet tripped:
- General solution: in "Magnet Systems"
window push Reset button
and wait for fault to clear (red rectangular is gone);
- If it did not help and if, according to the
"Magnet Systems"window,
it is Coil Voltage or Power Supply faults - call a Hall A
techical support on-call;
- If, on the camera display, Quench is gone red, you have to take
an access to the hall. In the hall on the power supply box, pull the lever,
wait for a couple minutes and then set it back;
Recommendation for data taking:
Stop data taking as soon as you saw magnet tripped.
- NMR control doesn't react and display shows all zeros:
NMR Teslameter communication failure:
- Push Reset in the "HRS Dipole
NMR" window;
- Re-enter all the settings, try to do it a couple of times, if it
does not work first;
- If everything fails, reboot;
Recommendation for data taking:
Unless you rebooted, no need to interrupt data taking. If you rebooted,
stop the run beforehand.
- All control displays show all zeros or
all nines: computer hang up:
Reboot, see above;
Recommendation for data taking:
Stop the run before you reboot.
Magnet Settings during E89033
| Hydrogen Kinematics |
Electron Arm
| Magnet | Current(A) | Field(T) |
| Q1 | - 1508.91 | 0.4958 |
| Q2 | - 862.47 | - 0.4509 |
| Q3 | - 797.43 | - 0.3970 |
| Dipole | 679.97 | 0.79266 |
|
|
Hadron Arm
| Magnet | Current(A) | Field(T) |
| Q1 | 745.58 | - |
| Q2 | 426.73 | - 0.2119 |
| Q3 | 394.55 | - 0.17632 |
| Dipole | 337.52 | 0.36795 |
|
|
| Oxygen Kinematics |
Electron Arm
| Magnet | Current(A) | Field(T) |
| Q1 | - 1496.43 | 0.49220 |
| Q2 | - 856.48 | - 0.44700 |
| Q3 | - 791.88 | - 0.39370 |
| Dipole | 675.18 | 0.78715 |
|
|
Hadron Arm
| Magnet | Current(A) | Field(T) |
| Q1 | 727.63 | - |
| Q2 | 416.46 | - 0.2068 |
| Q3 | 385.05 | - 0.17212 |
| Dipole | 329.56 | 0.35909 |
|
|
| Helium Levels and Lead Flow Limits |
Electron Arm
| Magnet | He Level(%) | Lead Flow(l/m) |
| Q1 | 45 | 60 |
| Q2 | 60 | 25 |
| Q3 | 60 | 25 |
| Dipole | 40 | 40 |
|
|
Hadron Arm
| Magnet | He Level(%) | Lead Flow(l/m) |
| Q1 | 45 | 60 |
| Q2 | 60 | 25 |
| Q3 | 60 | 25 |
| Dipole | 40 | 40 |
|
|
References
- OPS Manual (in a big binder in a counting house)
- J.LeRose, writeup on magnet operating
(in a separate binder in a counting house).
Go back to the
FPP Home Page
Please, send any comments on this page to
Sergey Malov
Last updated July 23,1997