March and April 2001 H(e,e'K) Runplan
3.395 GeV Preliminaries
Before Starting:
Before beginning check the beam conditions. The OTR should be in (if
possible). Fast feedback should also be on, both energy and orbit
locks and locked to Hall A. The 15 cm LH2 target should be in, and the
MCC should be told to setup for 75 uA (but that we want to start at 25
uA first).
Beam Energy:
Request a beam energy measurement ASAP. Ask Doug Higinbotham or Joe
Mitchell. The energy should be 3.395 GeV. If different by more
than 5 MeV (as measured by the Hall A arc-integral B*dl or the Hall A
ep), get an updated table of kinematics from P. Markowitz. If there is
any difficulty either with the energy or the locks, contact
A. Saha.
Trigger and DAQ: Page or
phone Bob Michaels. Ask him to: i) modify how we write data to the
storage silo, ii) change the timing of the kaon coincidence trigger.
Spectrometer Angles:
Page or phone the Hall A technician on call to come in and change the
angle. Phone MCC and request an access. Be sure to notify the
technician of the angle change (it may take two technicians to do this
angle change). Set the magnets second. Pay attention as the dipole
ramps up. The right-arm (e.g., electron for us) has been mis-behaving
but at higher momenta. Call John LeRose or Javier Gomez if there is
any difficulty. HALOG the magnet screen when settings are complete.<
Beam position: Less than
0.100 mm for x and y for 3A and 3B
Raster size:
2.0x2.0mm in SPOT or smaller.
Beam Current:
Start at 25 microA on the 15 cm LH2. Expect singles rates of 40 Khz at
25 microA.
Target:
15cm LH2
3.395 GeV First setting
Kinematics 17:
We will start at the Q2=1.90 (GeV/c)2, W=1.94 GeV,
low epsilon setting.
Spectrometer settings
| Spectrometer |
Momentum |
Angle |
Collimator |
| Right |
-0.8460 GeV/c |
-47.99° |
6msr |
| Left |
+1.986 GeV/c |
+12.50° |
6msr |
Data taking
If the angle
change is not yet underway as described above, change to 12.5 degrees
for the left (kaon) arm. Change the magnets. HALOG the magnet screen
when settings are complete.
Change the
electron angle simultaneously. When the technicians come out of the
hall, HALOG the angles and the vernier settings.
The trigger
setup needs to be set for a electron in the right arm and a kaon (of
momentum PK=1.986) in the left arm. Coincidence gate
width should be 80 ns.
The prescales for
T1,T2,T3 and T4 should be set to give less than 10 Hz trigger rate for
each. T5 should be set to 1. The T8 prescale should be set to
maximum. HALOG the prescale values.
Need to
check the header file: hdr_kaon17
The data
taking time is estimated to be 40 hours. Replay person (3rd) will need
to update the time by doing a simultaneous replay and keeping a
running sum on the white board of the number of kaons and the total
triggers. Also check detector spectra with the detector
kumac.
Take a
short (15 minute) run at 25 uA then increase current to 75 uA and take
the rest of the data.
Take a
short (30 minute) run at 25 uA or the maximum current on the 15 cm empty
target.
Put in the
optics 9-foil target and take a 15 minute mainly singles run.
Remember to change the prescales to 1 for T1 and T3. Keep T2 and T4
at less than 50 Hz. Use a beam current as high as possible keeping
deadtime less than 20%.
3.395 GeV Second setting
Kinematics 18:
The second setting is the Q2=2.35 (GeV/c)2, W=1.85 GeV,
low epsilon setting.
Spectrometer settings
| Spectrometer |
Momentum |
Angle |
Collimator |
| Right |
-0.7880 GeV/c |
-55.89° |
6msr |
| Left |
+1.893 GeV/c |
+12.50° |
6msr |
Data taking
Leave the kaon
angle as it is. Set the momenta to the new settings (-.7880 GeV/c and
+1.893 GeV/c). Both of these represent decreases in the momenta and
the magnets do NOT need to be cycled. Call John LeRose or Javier
Gomez if there is any difficulty. HALOG the magnet
screen.
While the magnets
are ramping, perform a harp scan and a beam current calibration. HALOG
the results.
Change the
electron angle simultaneously. When the angles have been reached,
type the results into the main GUI and HALOG the angles and the
vernier settings.
The trigger
setup should not need changed.
The prescales for
T1,T2,T3 and T4 should be set to give less than 10 Hz trigger rate for
each. T5 should be set to 1. The T8 prescale should be set to
maximum. HALOG the prescale values. The values from the previous
kinematics should be good.
Need to
check the header file : hdr_kaon18
The data
taking time is estimated to be 40 hours. Replay person (3rd) will need
to update the time by doing a simultaneous replay and keeping a
running sum on the white board of the number of kaons and the total
triggers. Also check detector spectra with the detector
kumac.
The current
should be 75 uAmps.
Take a
short (30 minute) run at 25 uA or the maximum current on the 15 cm empty
target.
Put in the
optics 9-foil target and take a 15 minute mainly singles run.
Remember to change the prescales to 1 for T1 and T3. Keep T2 and T4
at less than 50 Hz. Use a beam current as high as possible keeping
deadtime less than 20%.
3.395 GeV Third setting
Kinematics 19:
The third setting is the Q2=1.90 (GeV/c)2, W=1.905 GeV,
low epsilon setting.
Spectrometer settings
| Spectrometer |
Momentum |
Angle |
Collimator |
| Right |
-0.9279 GeV/c |
-45.70° |
6msr |
| Left |
+1.890 GeV/c |
+13.59° |
6msr |
Data taking
Set the momenta
to the new settings (-.9279 GeV/c and +1.890 GeV/c). The kaon arm is
still dropping in momenta and does not need cycled but the electronarm
now will need cycled. Call John LeRose or Javier Gomez if there is
any difficulty. HALOG the magnet screen.
While the magnets
are ramping, phone the Hall A technician on call and the MCC to get an
access for an angle change.
Change the
electron angle simultaneously. When the angles have been reached,
type the results into the main GUI and HALOG the angles and the
vernier settings.
The trigger
setup should not need changed.
The prescales for
T1,T2,T3 and T4 should be set to give less than 10 Hz trigger rate for
each. T5 should be set to 1. The T8 prescale should be set to
maximum. HALOG the prescale values. The values from the previous
kinematics should be good.
Need to
check the header file : hdr_kaon19
The data
taking time is estimated to be 40 hours. Replay person (3rd) will need
to update the time by doing a simultaneous replay and keeping a
running sum on the white board of the number of kaons and the total
triggers. Also check detector spectra with the detector
kumac.
The current
should be 75 uAmps.
Take a
short (30 minute) run at 25 uA or the maximum current on the 15 cm empty
target.
Put in the
optics 9-foil target and take a 15 minute mainly singles run.
Remember to change the prescales to 1 for T1 and T3. Keep T2 and T4
at less than 50 Hz. Use a beam current as high as possible keeping
deadtime less than 20%.
3.395 GeV Fourth setting
Kinematics 20:
The fourth 3.395 GeV setting is the Q2=2.35 (GeV/c)2, W=1.80 GeV,
low epsilon setting.
Spectrometer settings
| Spectrometer |
Momentum |
Angle |
Collimator |
| Right |
-0.8853 GeV/c |
-52.48° |
6msr |
| Left |
+1.741 GeV/c |
+13.81° |
6msr |
Data taking
Set the momenta
to the new settings (-.8853 GeV/c and +1.741 GeV/c). Both of these
represent decreases in the momenta and the magnets do NOT need to be
cycled. Call John LeRose or Javier Gomez if there is any
difficulty. HALOG the magnet screen.
While the magnets
are ramping, phone the Hall A technician on call and the MCC to get an
access for an angle change.
Change the
electron angle simultaneously. When the angles have been reached,
type the results into the main GUI and HALOG the angles and the
vernier settings.
The trigger
setup should not need changed.
The prescales for
T1,T2,T3 and T4 should be set to give less than 10 Hz trigger rate for
each. T5 should be set to 1. The T8 prescale should be set to
maximum. HALOG the prescale values. The values from the previous
kinematics should be good.
Need to
check the header file : hdr_kaon20
The data
taking time is estimated to be 40 hours. Replay person (3rd) will need
to update the time by doing a simultaneous replay and keeping a
running sum on the white board of the number of kaons and the total
triggers. Also check detector spectra with the detector
kumac.
The current
should be 75 uAmps.
Take a
short (30 minute) run at 25 uA or the maximum current on the 15 cm empty
target.
Put in the
optics 9-foil target and take a 15 minute mainly singles run.
Remember to change the prescales to 1 for T1 and T3. Keep T2 and T4
at less than 50 Hz. Use a beam current as high as possible keeping
deadtime less than 20%.
4.232 GeV Preliminaries
Before Starting:
Before beginning check the beam conditions. The OTR should be in (if
possible). Fast feedback should also be on, both energy and orbit
locks and locked to Hall A. The 15 cm LH2 target should be in, and the
MCC should be told to setup for 75 uA (but that we want to start at 25
uA first).
Beam Energy:
Request a beam energy measurement ASAP. Ask Doug Higinbotham or Joe
Mitchell. The energy should be 4.232 GeV. If different by more
than 5 MeV (as measured by the Hall A arc-integral B*dl or the Hall A
ep), get an updated table of kinematics from P. Markowitz. If there is
any difficulty either with the energy or the locks, contact
A. Saha.
Spectrometer Angles:
Page or phone the Hall A technician on call to come in and change the
angle. Phone MCC and request an access. Be sure to notify the
technician of the angle change (it may take two technicians to do this
angle change). Set the magnets second. Pay attention as the dipole
ramps up. The right-arm (e.g., electron for us) has been mis-behaving
but at higher momenta. Call John LeRose or Javier Gomez if there is
any difficulty. HALOG the magnet screen when settings are complete.<
Beam position: Less than
0.100 mm for x and y for 3A and 3B
Raster size:
2.0x2.0mm in SPOT or smaller.
Beam Current:
Start at 25 microA on the 15 cm LH2. Expect singles rates of 40 Khz at
25 microA.
Target:
15cm LH2
4.232 GeV First setting
Kinematics 21:
We will start at the Q2=1.90 (GeV/c)2, W=2.14 GeV,
low epsilon setting.
Spectrometer settings
| Spectrometer |
Momentum |
Angle |
Collimator |
| Right |
-1.224 GeV/c |
-34.90° |
6msr |
| Left |
+2.521 GeV/c |
+12.53° |
6msr |
Data taking
If the angle
change is not yet underway as described above, change to 12.53 degrees
for the left (kaon) arm. Change the magnets. HALOG the magnet screen
when settings are complete.
Change the
electron angle simultaneously. When the technicians come out of the
hall, HALOG the angles and the vernier settings.
The trigger
setup needs to be set for a electron in the right arm and a kaon (of
momentum PK=2.521) in the left arm. Coincidence gate
width should be 80 ns.
The prescales for
T1,T2,T3 and T4 should be set to give less than 10 Hz trigger rate for
each. T5 should be set to 1. The T8 prescale should be set to
maximum. HALOG the prescale values.
Need to
check the header file: hdr_kaon21
The data
taking time is estimated to be 20 hours. Replay person (3rd) will need
to update the time by doing a simultaneous replay and keeping a
running sum on the white board of the number of kaons and the total
triggers. Also check detector spectra with the detector
kumac.
Take a
short (15 minute) run at 25 uA then increase current to 75 uA and take
the rest of the data.
Take a
short (30 minute) run at 25 uA or the maximum current on the 15 cm empty
target.
Put in the
optics 9-foil target and take a 15 minute mainly singles run.
Remember to change the prescales to 1 for T1 and T3. Keep T2 and T4
at less than 50 Hz. Use a beam current as high as possible keeping
deadtime less than 20%.
4.232 GeV Second setting
Kinematics 22:
The second 4.232 GeV setting is the Q2=2.35 (GeV/c)2, W=2.08 GeV,
low epsilon setting.
Spectrometer settings
| Spectrometer |
Momentum |
Angle |
Collimator |
| Right |
-1.120 GeV/c |
-40.79° |
6msr |
| Left |
+2.542 GeV/c |
+12.53° |
6msr |
Data taking
Leave the kaon
angle as it is. Set the momenta to the new settings (-1.143 GeV/c and
+2.542 GeV/c). Only the kaon arm needs to be cycled.
Call John LeRose or Javier
Gomez if there is any difficulty. HALOG the magnet
screen.
While the magnets
are ramping, perform a harp scan and a beam current calibration. HALOG
the results.
Change the
electron angle simultaneously. When the angles have been reached,
type the results into the main GUI and HALOG the angles and the
vernier settings.
The trigger
setup should not need changed.
The prescales for
T1,T2,T3 and T4 should be set to give less than 10 Hz trigger rate for
each. T5 should be set to 1. The T8 prescale should be set to
maximum. HALOG the prescale values. The values from the previous
kinematics should be good.
Need to
check the header file : hdr_kaon22
The data
taking time is estimated to be 40 hours. Replay person (3rd) will need
to update the time by doing a simultaneous replay and keeping a
running sum on the white board of the number of kaons and the total
triggers. Also check detector spectra with the detector
kumac.
The current
should be 75 uAmps.
Call the Hall A technicians to
prepare us to go OUT from 12.53 degrees. Take a
short (30 minute) run at 25 uA or the maximum current on the 15 cm empty
target.
Put in the
optics 9-foil target and take a 15 minute mainly singles run.
Remember to change the prescales to 1 for T1 and T3. Keep T2 and T4
at less than 50 Hz. Use a beam current as high as possible keeping
deadtime less than 20%.
4.232 GeV NEW Third setting
Kinematics 25:
We are changing the order in case the energy per pass is changed. The next
setting is the Q2=2.35 (GeV/c)2, W=1.98 GeV, low
epsilon setting.
Spectrometer settings
| Spectrometer |
Momentum |
Angle |
Collimator |
| Right |
-1.333 GeV/c |
-37.27° |
6msr |
| Left |
+2.266 GeV/c |
+14.65° |
6msr |
Data taking
Set the momenta
to the new settings (-1.360 GeV/c and +2.266 GeV/c). Cycle the kaon
arm. Call John LeRose or Javier Gomez if there is any
difficulty. HALOG the magnet screen.
Change the
electron angle simultaneously. When the angles have been reached,
type the results into the main GUI and HALOG the angles and the
vernier settings.
The trigger
setup should not need changed.
The prescales for
T1,T2,T3 and T4 should be set to give less than 10 Hz trigger rate for
each. T5 should be set to 1. The T8 prescale should be set to
maximum. HALOG the prescale values. The values from the previous
kinematics should be good.
Need to
check the header file : hdr_kaon25
The data
taking time is estimated to be 40 hours. Replay person (3rd) will need
to update the time by doing a simultaneous replay and keeping a
running sum on the white board of the number of kaons and the total
triggers. Also check detector spectra with the detector
kumac.
The current
should be 75 uAmps.
Take a
short (30 minute) run at 25 uA or the maximum current on the 15 cm empty
target.
Put in the
optics 9-foil target and take a 15 minute mainly singles run.
Remember to change the prescales to 1 for T1 and T3. Keep T2 and T4
at less than 50 Hz. Use a beam current as high as possible keeping
deadtime less than 20%.
4.232 GeV NEW Fourth setting
Kinematics 28:
The next setting is the Q2=1.90 (GeV/c)2, W=2.00 GeV,
low epsilon setting.
Spectrometer settings
| Spectrometer |
Momentum |
Angle |
Collimator |
| Right |
-1.526 GeV/c |
-31.15° |
6msr |
| Left |
+2.147 GeV/c |
+15.52° |
6msr |
Data taking
Set the momenta
to the new settings (-1.557 GeV/c and +2.147 GeV/c). Cycle the kaon
arm. Call John LeRose or Javier Gomez if there is any
difficulty. HALOG the magnet screen.
Change the
electron angle simultaneously. When the angles have been reached,
type the results into the main GUI and HALOG the angles and the
vernier settings.
The trigger
setup should not need changed.
The prescales for
T1,T2,T3 and T4 should be set to give less than 10 Hz trigger rate for
each. T5 should be set to 1. The T8 prescale should be set to
maximum. HALOG the prescale values. The values from the previous
kinematics should be good.
Need to
check the header file : hdr_kaon28
The data
taking time is estimated to be 20 hours. Replay person (3rd) will need
to update the time by doing a simultaneous replay and keeping a
running sum on the white board of the number of kaons and the total
triggers. Also check detector spectra with the detector
kumac.
The current
should be 75 uAmps.
Take a
short (30 minute) run at 25 uA or the maximum current on the 15 cm empty
target.
Put in the
optics 9-foil target and take a 15 minute mainly singles run.
Remember to change the prescales to 1 for T1 and T3. Keep T2 and T4
at less than 50 Hz. Use a beam current as high as possible keeping
deadtime less than 20%.
4.232 GeV NEW Fifth setting
Kinematics 23:
The third 4.232 GeV setting is the Q2=2.35 (GeV/c)2, W=1.80 GeV,
low epsilon setting.
Spectrometer settings
| Spectrometer |
Momentum |
Angle |
Collimator |
| Right |
-1.688 GeV/c |
-32.99° |
6msr |
| Left |
+1.741 GeV/c |
+18.59° |
6msr |
Data taking
Get the technicians to take us to
the new kaon angle. Set the momenta
to the new settings (-1.722 GeV/c and +1.741 GeV/c). The kaon arm is
still dropping in momenta and does not need cycled but the electron arm
now will need cycled. Call John LeRose or Javier Gomez if there is
any difficulty. HALOG the magnet screen.
While the magnets
are ramping, phone the Hall A technician on call and the MCC to get an
access for an angle change.
Change the
electron angle simultaneously. When the angles have been reached,
type the results into the main GUI and HALOG the angles and the
vernier settings.
The trigger
setup should not need changed.
The prescales for
T1,T2,T3 and T4 should be set to give less than 10 Hz trigger rate for
each. T5 should be set to 1. The T8 prescale should be set to
maximum. HALOG the prescale values. The values from the previous
kinematics should be good.
Need to
check the header file : hdr_kaon23
The data
taking time is estimated to be 20 hours. Replay person (3rd) will need
to update the time by doing a simultaneous replay and keeping a
running sum on the white board of the number of kaons and the total
triggers. Also check detector spectra with the detector
kumac.
The current
should be 75 uAmps.
Take a
short (30 minute) run at 25 uA or the maximum current on the 15 cm empty
target.
Put in the
optics 9-foil target and take a 15 minute mainly singles run.
Remember to change the prescales to 1 for T1 and T3. Keep T2 and T4
at less than 50 Hz. Use a beam current as high as possible keeping
deadtime less than 20%.
4.232 GeV NEW Sixth setting
Kinematics 24:
The next setting is the Q2=2.35 (GeV/c)2, W=1.85 GeV,
low epsilon setting.
Spectrometer settings
| Spectrometer |
Momentum |
Angle |
Collimator |
| Right |
-1.593 GeV/c |
-33.99° |
6msr |
| Left |
+1.893 GeV/c |
+17.48° |
6msr |
Data taking
Set the momenta
to the new settings (-1.625 GeV/c and +1.893 GeV/c). Cycle the kaon
arm. Call John LeRose or Javier Gomez if there is any
difficulty. HALOG the magnet screen.
Change the
electron angle simultaneously. When the angles have been reached,
type the results into the main GUI and HALOG the angles and the
vernier settings.
The trigger
setup should not need changed.
The prescales for
T1,T2,T3 and T4 should be set to give less than 10 Hz trigger rate for
each. T5 should be set to 1. The T8 prescale should be set to
maximum. HALOG the prescale values. The values from the previous
kinematics should be good.
Need to
check the header file : hdr_kaon24
The data
taking time is estimated to be 20 hours. Replay person (3rd) will need
to update the time by doing a simultaneous replay and keeping a
running sum on the white board of the number of kaons and the total
triggers. Also check detector spectra with the detector
kumac.
The current
should be 75 uAmps.
Take a
short (30 minute) run at 25 uA or the maximum current on the 15 cm empty
target.
Put in the
optics 9-foil target and take a 15 minute mainly singles run.
Remember to change the prescales to 1 for T1 and T3. Keep T2 and T4
at less than 50 Hz. Use a beam current as high as possible keeping
deadtime less than 20%.
4.232 GeV NEW Seventh setting
Kinematics 26:
The next setting is the Q2=1.90 (GeV/c)2, W=1.905 GeV,
low epsilon setting.
Spectrometer settings
| Spectrometer |
Momentum |
Angle |
Collimator |
| Right |
-1.730 GeV/c |
-29.21° |
6msr |
| Left |
+1.890 GeV/c |
+17.75° |
6msr |
Data taking
Set the momenta
to the new settings (-1.765 GeV/c and +1.890 GeV/c). Cycle the electron
arm. Call John LeRose or Javier Gomez if there is any
difficulty. HALOG the magnet screen.
Change the
electron angle simultaneously. When the angles have been reached,
type the results into the main GUI and HALOG the angles and the
vernier settings.
The trigger
setup should not need changed.
The prescales for
T1,T2,T3 and T4 should be set to give less than 10 Hz trigger rate for
each. T5 should be set to 1. The T8 prescale should be set to
maximum. HALOG the prescale values. The values from the previous
kinematics should be good.
Need to
check the header file : hdr_kaon26
The data
taking time is estimated to be 20 hours. Replay person (3rd) will need
to update the time by doing a simultaneous replay and keeping a
running sum on the white board of the number of kaons and the total
triggers. Also check detector spectra with the detector
kumac.
The current
should be 75 uAmps.
Take a
short (30 minute) run at 25 uA or the maximum current on the 15 cm empty
target.
Put in the
optics 9-foil target and take a 15 minute mainly singles run.
Remember to change the prescales to 1 for T1 and T3. Keep T2 and T4
at less than 50 Hz. Use a beam current as high as possible keeping
deadtime less than 20%.
4.232 GeV NEW Eighth setting
Kinematics 27:
The next setting is the Q2=1.90 (GeV/c)2, W=1.94 GeV,
low epsilon setting.
Spectrometer settings
| Spectrometer |
Momentum |
Angle |
Collimator |
| Right |
-1.650 GeV/c |
-29.93° |
6msr |
| Left |
+1.986 GeV/c |
+16.85° |
6msr |
Data taking
Set the momenta
to the new settings (-1.683 GeV/c and +1.986 GeV/c). Cycle the electron
arm. Call John LeRose or Javier Gomez if there is any
difficulty. HALOG the magnet screen.
Change the
electron angle simultaneously. When the angles have been reached,
type the results into the main GUI and HALOG the angles and the
vernier settings.
The trigger
setup should not need changed.
The prescales for
T1,T2,T3 and T4 should be set to give less than 10 Hz trigger rate for
each. T5 should be set to 1. The T8 prescale should be set to
maximum. HALOG the prescale values. The values from the previous
kinematics should be good.
Need to
check the header file : hdr_kaon27
The data
taking time is estimated to be 20 hours. Replay person (3rd) will need
to update the time by doing a simultaneous replay and keeping a
running sum on the white board of the number of kaons and the total
triggers. Also check detector spectra with the detector
kumac.
The current
should be 75 uAmps.
Take a
short (30 minute) run at 25 uA or the maximum current on the 15 cm empty
target.
Put in the
optics 9-foil target and take a 15 minute mainly singles run.
Remember to change the prescales to 1 for T1 and T3. Keep T2 and T4
at less than 50 Hz. Use a beam current as high as possible keeping
deadtime less than 20%.
4.232 GeV Ninth and Tenth settings
Kinematics 29 and
30: The twenty-ninth and
thirtieth settings are the LT-interfence part. Call in the Hall A "tech
on call" halfway through the twenty-ninth measurement to prepare for the
small angle. Here the electron will stay fixed and the hadron
momentum will be constant between 29th and 30th settings.
Settings: Change the
electron arm first then take a 5 minute single arm run with the
multifoil targets. Then change the hadron arm and take another 5
minute single arm run with the multifoil target.
Beam and Target:
Double check that fast feedback locks on both energy and orbit are on.
Spot should give a raster of 2.0x2.0mm or smaller. Change to the LH2
target. Again start at 25 microA on the 15 cm LH2.
Spectrometer settings:
| Spectrometer |
Momentum |
Angle |
Collimator |
| Right |
-1.650 GeV/c |
-29.93° |
6msr |
| Left |
+1.930 GeV/c |
+21.18/+12.51° |
6msr |
Data taking
Change the
prescales back to what they were (as a starting point). Be sure the
trigger is properly loaded (it should be the same).
HALOG the settings (magnets, angles, prescales).
The prescales for T1,T2,T3
and T4 should be set to give less than 10 Hz trigger rate for each. T5
should be set to 1. The T8 prescale should be set to maximum. HALOG the
prescale values.
Need to
modify the header file : hdr_kaon29,
hdr_kaon30 appropriately.
The data
taking time is estimated to be 4.0 hours each for settings.
Will need to update by doing a simultaneous
replay.
Take a
short (15 minute) run at 25 uA then increase current to 75 uA and take
the rest of the data.
Take a
short (15 minute) run at 25 uA on the 15 cm empty
target for each setting.
Change to
angle 30, take the hydrogen data (again 4 hours) then switch to the
multifoil target and take 5 minutes at 10 uA.