Moller Measurements for E06-010 Transversity, Nov 2008 - Feb 2009
The Moller results are presented in a table below and on a plot 1
and plot 2.
# date day* Moller Moller with HALLA:p Syst.err.
---------------------------------------------------------------------
1 11.02.2008 12 +88.41 +/- 0.22 +88.41 +/- 0.22 +/- 2.0%
2 11.12.2008 22 -74.94 +/- 0.14 -79.64 +/- 0.14 +/- 3.0%
3 11.17.2008 27 -74.46 +/- 0.11 -80.06 +/- 0.11 +/- 3.0%
4 11.20.2008 30 -74.22 +/- 0.17 -77.08 +/- 0.17 +/- 3.0%
5 11.28.2008 38 +80.97 +/- 0.19 +79.64 +/- 0.19 +/- 3.0%
6 12.08.2008 48 -54.16 +/- 0.20 +54.39 +/- 0.20 +/- 3.0%
7 12.09.2008 49 -79.16 +/- 0.16 +79.76 +/- 0.16 +/- 3.0%
8 12.17.2008 57 -80.00 +/- 0.18 -78.86 +/- 0.16 +/- 3.0%
9 01.13.2009 63 -81.12 +/- 0.15 -79.88 +/- 0.16 +/- 2.0%
10 01.24.2009 74 -79.01 +/- 0.13 -79.88 +/- 0.13 +/- 2.0%
11 02.03.2009 85 -75.84 +/- 0.16 -79.11 +/- 0.16 +/- 2.0%
* day - day of experiment running
Measurement 1: (11.02.2008)
There is no energy fluctuation. Beam energy correction was not done.
Measurements 2 - 8
The Moller measurements 2-8 were done at ~0.5uA curent. At this current HALLA:p energy reading is more noisy and for a part of
measurements energy correction can not be done at all. This is the reason to increase systematic error from 2% to 3%.
Measurement 6: (08.12.2008):
From Eugine's words the beam polarization was low due to the injector problem.
Measurement 9: Spin Dance (01.13.2009)
The plot is given here .
The details of the measurements can be found here .