Moller Measurements in Jul-Nov 1999
The longitudinal beam polarization is calculated as:
P=A/Ptarg/cos(a)/Azz,
where
The laser phase had been checked before the measurements.
The table of all the measurements is shown here:
# m d h 1-st run GeV Polarization COS(spin) uA Laser:power atten slit loss 1/2wave B C 1 7 10 10 7006 3.3250 75.63 +/- 0.18 -0.997 0.503 73.0 500. 15.90 0.66 0 0 0 2 7 12 17 7023 3.3250 -74.35 +/- 0.17 -0.997 0.545 73.0 500. 15.80 0.81 1 1 0 3 7 12 18 7053 3.3250 -74.88 +/- 0.20 -0.997 0.137 73.0 500. 13.90 0.44 1 1 0 4 7 13 13 7066 3.3250 -76.02 +/- 0.18 1.000 0.509 0.0 0. 0.00 0.00 0 1 1 5 7 15 14 7127 3.3250 76.27 +/- 0.19 -1.000 0.478 73.0 500. 15.50 0.99 0 0 0 6 7 15 20 7147 3.3250 76.56 +/- 0.18 -1.000 0.547 73.0 500. 16.00 0.64 0 0 0 7 7 15 21 7177 3.3250 76.37 +/- 0.18 -1.000 0.231 73.0 500. 14.50 0.53 0 1 0 8 7 22 8 7190 3.3250 -71.37 +/- 0.17 -0.997 0.575 73.0 500. 15.75 0.91 1 1 1 9 8 4 0 7318 1.1480 79.72 +/- 0.26 0.983 0.189 73.0 500. 14.58 0.45 0 1 1 10 8 12 12 7393 0.8570 78.53 +/- 0.21 -0.997 0.206 73.0 200. 16.50 0.49 0 0 1 11 8 14 17 7424 1.6690 77.57 +/- 0.19 -0.920 0.275 73.0 355. 15.50 0.78 0 1 1 12 8 18 12 7452 1.6690 77.54 +/- 0.19 -0.920 0.253 73.0 328. 16.15 0.38 0 1 1 13 8 19 21 7477 4.1080 -80.73 +/- 0.14 0.996 0.525 73.0 500. 15.70 0.88 0 1 1 14 8 24 7 7608 4.1080 -80.17 +/- 0.15 0.995 0.303 73.0 600. 15.45 0.66 0 1 1 15 8 24 18 7639 1.6690 74.85 +/- 0.16 -0.921 0.301 73.0 500. 15.30 0.73 0 1 1 16 8 26 16 7664 1.6690 75.87 +/- 0.17 -0.919 0.271 73.0 500. 15.00 0.72 0 0 1 17 8 27 19 7690 1.6690 75.73 +/- 0.16 -0.920 0.290 73.0 500. 15.01 0.77 0 1 1 18 8 30 10 7718 1.6690 76.11 +/- 0.16 -0.920 0.291 73.0 500. 15.30 0.71 0 1 0 19 8 31 15 7743 1.6690 75.94 +/- 0.17 -0.920 0.255 73.0 500. 15.20 0.65 0 1 1 20 9 21 9 7778 1.2680 65.08 +/- 0.20 -0.976 0.220 75.0 200. 17.10 0.37 0 1 1 21 9 24 10 7807 2.5020 64.62 +/- 0.18 -0.988 0.308 75.0 600. 16.20 0.29 0 1 0 22 9 27 11 7834 2.5020 62.90 +/- 0.15 -0.988 0.453 76.0 600. 16.20 0.43 0 1 1 23 10 2 12 7863 2.5010 61.07 +/- 0.20 -0.988 0.401 76.0 600. 15.46 0.87 0 1 1 24 10 5 16 7915 2.5010 58.75 +/- 0.21 -0.988 0.362 76.0 600. 15.00 0.96 0 0 1 25 10 8 13 7947 2.5010 66.55 +/- 0.20 -0.988 0.284 76.0 600. 15.00 0.75 0 1 1 26 10 15 17 7995 2.4980 68.36 +/- 0.22 -0.988 0.324 76.0 600. 16.20 0.30 0 1 0 27 10 20 15 8016 3.1150 -68.71 +/- 0.23 0.995 0.305 76.0 430. 16.70 0.20 0 1 0 28 10 23 9 8049 3.1150 -67.70 +/- 0.22 1.000 0.293 76.0 600. 17.40 0.14 0 1 1 29 10 29 14 8090 1.9470 -70.36 +/- 0.17 1.000 0.249 75.0 600. 15.30 0.61 0 1 0 30 11 2 13 8118 1.9470 -73.28 +/- 0.17 1.000 0.296 75.0 600. 14.80 0.75 0 0 0 31 11 5 18 8138 4.7920 -70.00 +/- 0.18 0.997 0.305 75.0 600. 15.15 0.80 0 1 0 32 11 6 15 8167 1.0000 -73.98 +/- 0.17 0.994 0.197 75.0 600. 14.80 0.50 0 1 0where the date is given in the columns 2-4, the average polarization for 2 target angles is presented in the columns called "Polarization". The 1-st Moller run of the measurement is shown in column 4. The column named "COS(spin)" shows the value of cosine of the expected spin angle in the hall. The column named "loss" shows the estimated value of the "additional" attenuation parameter (see the details here). The columns named "B" and "C" show whether the lasers of Halls B and C were on or off during the measurements.
The Moller results, along with the Mott polarimeter results and indications of the injector and other changes are presented on a plot. The PAW macro which creates this plot is available from E.Chudakov.
The line "radiation damage" indicate the moment when the Moller detector was slightly damaged by radiation during the tune-up of the beam. Mostly, the left top lead glass counter was damaged, loosing about 40% of the amplitude. In several days the detector has practically recovered.