status of Forward Tracker 4/15/2004



1.) Simulation:


3.) Readout:



5.) Test Apparatus
Student Jeremy Dobbins has return from CERN after working with Fabio Sauli. While at CERN he built a second Qweak detector proto-type. This proto-type has 3 GEM premplifier stages and a trigger output line from the last GEM foil.
Second Qweak Prototype.
The gain as a function of GEM foil potential difference has been measured using 8 keV X-ray and 2 different gas mixtures. The 75% Ar and 25 CO2 mixture is a standard mixture while the 70/30 mixture is a more expensive special order. At present it is $25 for the 75/25 and $150 for the 70/30 mixture.
The detector's response in a typical 2 cm x 2cm region was measured. The detector output fluctuates by up to 10% for the 8 keV collimated X-ray dose. Summing 16 readout lines and amplifying by a factor of 5 results in a 450-500 mV output (so one unamplified line would be around 5 mV into the Helix board)


Tony Forest