Appendix D   LSOP for CLAS Calibration Lasers

D.1   Detectors and Environment

The the highest level of the forward carriage in Hall B at Jefferson Lab contains two identical class IIIb lasers used for calibrating equipment for the CLAS detector. The lasers are used to provide light sources for photomultiplier tubes of the CLAS Time-of-Flight (TOF) and calorimeter systems. The locations of the lasers on the Forward Carriage are shown schematically in Figure 1. Three additional lasers for calibration of the TOF scintillators are located on the north Clam Shell, south Clam Shell and Space Frame. All laser setups are similar, so the same precautions must be taken for each.This Laser Standard Operating Procedure (LSOP) gives a description of the lasers, the hazards present, requirements for personnel and procedures for operation.

The distribution of fibers for both systems is shown in Figure 2.

D.2   Personnel

The lasers may only be operated by personnel which have

CEBAF personnel or outside visitors, who have not completed the required training, may enter the laser control area under the following conditions:

or when

D.3   Laser

Table 1 lists the characteristics of the Laser Photonics LN203C Nitrogen laser.

TABLE 7. Specifications for Photonics LN203C Nitrogen laser

Specification

LN203C

Spectral Output

337.1 nm

Spectral Bandwidth

0.1 nm

Pulsewith (FWHM)

600 ps

Energy/pulse

100 mJ

Energy Stability

ü 3%

Maximum Average Power

5.0 mW

Maximum Repetition Rate

50 Hz

Beam Dimensions (hor x ver)

5.5 x 3.1 mm2

Beam Divergence (hor x ver, half angle)

6.2 x 2.5 mrad2

Gas Consumption (99.995\% pure N$_2$)

1 l/min @ 10 Hz

Dimensions

28.5 x 8.5 x 5.25 in3

Weight

20 lbs

D.4   Hazards

1. Laser Beam Hazards

Class 1 eyesafe laser conditions exist except during alignment procedures. During alignment procedures, the primary beam hazard associated with class IIIb lasers consists of eye injury caused by viewing the beam either directly or by specular reflection. 99% of the laser light (337 nm) entering the eye is absorbed in the cornea, aqueous and lens. The primary adverse effect from accidental viewing is the production of cataracts. To minimize the laser beam hazard associated with alignment procedures,

  • supplied laser safety goggles must be worn.
  • laser shroud must be placed between the access to the work area and the laser beam. The laser is not a fire hazard, so any UV absorbing material would suffice. A dark cloth is used to cover the area where laser light could escape.
2. Non-beam hazards

The following non-beam hazards exist:

  • trip/fall: access to the CLAS nitrogen lasers is limited to an extension ladder. To minimize hazards associated with the use of the ladder:
  • The ladder must be tied off securely

    All personnel must wear safety shoes

    One person on the ladder at a time

    Nothing is to carried up the ladder. Anything that is to be brought up to the laser work area must be lifted up. Small items can be lifted with a rope.

  • bump hazard: To minimize the bump hazard associated with the use of the extension ladder, post danger sign on the ladder noting the hazard and the requirement for a hard hat.
  • electric shock: If interlocks are bypassed and high voltage is exposed, there is potential for electric shock. No interlocks are to be bypassed, except by qualified service representatives.
  • Fire: a portable fire extinguisher is located in the work area
  • Lighting: Most of the emergency lighting for the Hall is located at lower levels and would not be of benefit in the event of a power outage; a standby flashlight is needed.
  • Sudden release of energy from cylinder: N2 cylinder must be secured.

D.5   Laser Environment

The location of the LN203C nitrogen lasers is shown in Figure 1. Beam height is approximately 3' above the floor. During normal operation, the laser beam is extracted via a fiber bundle array 1 . These measures prevent unintentional interception of the beam, either directly or by specular reflection. Under special circumstances, detailed below, the laser beam is exposed, and precautionary measures must be taken. The laser light is confined to the Rf-shielding box except during alignment. During alignment, the light is confined by draping a laser shroud over the end of the plexiglass fiber-optic box. Before laser work begins, personnel should make sure that they have a clear path to an exit.

D.6   Remote Control and Operation

The laser can be remotely controlled during routine operation by using specifically designed circuits 2 for the laser. The remote functions including power, pulse rate, enable and disable functions which can be used to automatically shut down operation in case of emergency. An emission time delay is built into the remote control: the yellow beacon energizes when the laser power is turned on, and lasing is delayed by 5 minutes. Once the laser yellow beacon is energized, goggles must be worn.

D.7   Procedures

In this section we review the various procedures that are required to operate the laser. Hazards are least likely to occur during normal operation. During maintenance and alignment, the laser box is open and beam and non-beam hazards are present. The manual for the LN203C laser details installation (Section III), normal operation (Section IV), maintenance (Section V) and trouble shooting (Section VI) of the laser.

When working with the lasers, the personnel must not wear jewelry or clothing which presents a specular reflection hazard. Protective eye wear must be worn with a minimum optical density OD = 1.4 at 337 nm (see Section 7 for the required eye wear). Safety UV-absorbing glasses that meet this requirement are available in the immediate area.

D.8   Maintenance and Initial Alignment

In this case the rf-shielding box is open, the on/off switch is activated, the N2 gas flow rate is adjusted, and the warm-up period initiated. To prevent optical hazards the beam stop should remain closed for this operation except during the initial alignment period. Standard electrical hazard prevention procedures should be followed. To minimize optical hazards, no one will be allowed in the laser control area except trained and qualified personnel who are wearing UV-absorbing safety glasses (see Section 7).

D.9   Scintillator Calibration (Class 1 Conditions)

The laser beams are directed to fiber bundle arrays at the exit of the RF-shielding boxes. The fibers in the bundle are connected to the TOF scintillators or directed to scintillators on the calorimeters Under these circumstances, the lasers shall be considered to pose a Class 1 hazard, as no portion of the beam is exposed.

D.10   Controls

The laser control area will be posted with a danger sign describing the tests being conducted. Figure 1 will be also be posted to indicate the laser control area. The key which turns the laser on shall be kept separate when the laser is not in use, to avoid unintentionally enabling operation. Eyewear is provided and shall be used during maintenance, alignment and any activity which allows viewing of the laser beam. A high voltage safety cover interlock is located on the rear panel of the laser. Removal of the top access panel is required for servicing and will automatically break the interlock. Any interlock interruption will require the high voltage enable switch to be re-engaged to resume operation.

D.11   MPE Calculations

We calculate the optical density, D, of protective eye wear required for laser operation, assuming the laser is continuous. The optical density is given in terms of the radiant exposure, H, and maximum permissible exposure, MPE, by

 

 

 

where a is the limiting aperture (1 mm), f is the maximum operating frequency (50 Hz), Ep is the maximum energy per pulse (100 mJ) and N is the number of pulses (frequency times observation time). With these parameters we calculate that

For an observation time of 10 s, N = 500, D = 0.8. For an observation time of 1 minute, N = 3000, D = 1.4.

Next, an MPE was calculated using the pulse width as exposure time by Dr. John Leonowich, a consultant for CEBAF. 3 This calculation yielded a required O.D. of 1.23. We use the more conservative estimate of 1.4.

D.12   Authorized Personnel

The following personnel are trained and qualified to operate the lasers in Hall B:

 

TABLE 8. List of Authorized Personnel. Their signature indicates
that they have read and understood the TOSP for laser operation.

Subsytem

Personnel

Signature

Date

TOF

Elton Smith

 

 

 

Wooyoung Kim

 

 

 

Simon Taylor

 

 

 

Kijun Park

 

 

 

Pyunghun Kim

 

 

 

Byungjoon Lee

 

 

 

 

 

 

Calorimeter

Kevin Giovanetti

 

 

 

Young Soo Sohn

 

 

 

Donghee Kim

 

 

 

Wonha Ko

 

 

 

Minjeong Kim

 

 

 

Chanhoon Chung

 

 

 

Minsuk Kim

 

 

No other personnel shall be added to the above list without authorization by Elton Smith.


1. Original calculations were made for a single 600 mm fiber. These computations become conservative for the present situation, where the laser beam is divided among the fibers in a fiber bundle. Measurements show that a single fiber transmits much less than 1.3% of the total. The intensity from a single fiber was unmeasurable with the power meter.

2. Circuit design by Kevin Giovanetti and Rob Atkins, James Madison University. The layout and implementation of the circuit is by Randy Wojcik, Detector Group.

3. John Leonowich, Battelle Laboratories engineering worksheet, Feb 2, 1995.